10011896-001
8
To install new batteries:
1. Stop the pump by pressing
and holding
⁄
.
2. When you see 3 sets of dashes
appear, release
⁄
.
3. Push down and hold the arrow
button on the battery door while
sliding the door off.
4. Remove the used batteries. Pulling
on the end of the battery strap will
make battery removal easier.
5. Match the
+
and
–
markings on the
new batteries with the markings
inside the battery compartment.
Insert the batteries over the top of
the battery strap. The pump will
beep if the batteries are inserted
correctly.
6. Replace the battery door. The pump
will power up automatically.
NOTES:
• If you put the batteries in backwards, the display will remain
blank. Reinsert the batteries, making sure to match the
+
and
–
markings.
• The power-up sequence will start, the pump will go through
an electronic self-test, and the pump will beep 6 times at the
end of the power-up sequence. All of the display indicators, the
software revision level, and each parameter will appear briefly.
STOPPED
Stopping
---- ---- ----
10011896-001.indd 8
2/27/2018 7:08:12 PM