10011896-001
15
3. Place the pump upright on a firm,
flat surface. Press down so the
cassette fits tightly against the
pump.
4. Insert a coin into the lock, push
in, and turn counterclockwise
until the line on the lock lines
up with the arrow on the side
of the pump and you feel the
button click into place.
WARNING: Attach the cassette (the part of the CADD
™
medication cassette
reservoir or CADD® administration
set that attaches to the pump) properly. An improperly
attached or detached cassette could result in unregulated
gravity infusion of medication from the fluid container or a
reflux of blood, which could result in death or serious injury.
5. Gently twist, push, and pull on
the cassette to make sure it is
firmly attached.