Chapter 4: Operating Instructions
4-10
Oximeter Clinician’s Operation Manual
•
For “Y” sensors, wrap sensors, and disposable sensors: Align the
sensor’s red light with the detector so they are less than 1/8 inch
away from each other. Make sure the SENSOR indicator is flash-
ing on the oximeter.
•
For the finger sensor and ear sensor: Make sure the SENSOR indi-
cator is flashing on the oximeter.
7. You are now ready to attach the
sensor to the patient.
Figure 4-6: Attach the
finger sensor.
Attach the finger sensor to the
patient as shown. Be sure to
fully insert the patient’s finger
into the sensor. For patients
with long fingernails, use the
universal “Y” sensor.
Cleaning or Disinfecting the Sensors
Clean or disinfect reusable sensors before attaching to a new patient.
2
WARNING! Do not autoclave, ethylene oxide sterilize, or
immerse the sensors in liquid.
CAUTION! Unplug the sensor from the monitor before cleaning
or disinfecting.
Clean the sensor with a soft cloth moistened in water or a mild soap
solution. To disinfect the sensor, wipe the sensor with isopropyl alcohol.