Oximeter
Parameters Monitored
SpO
2
Pulse Rate
Pulse Strength
Displays, Indicators, & Keys SpO
2
3-digit LED display, 0.43 inches
(10.9 mm) high.
Pulse Rate
3-digit LED display, 0.375 inches
(9.5 mm) high.
Pulse
Strength
Logarithmically scaled 8-seg-
ment LED bar graph.
SENSOR
Sensor alert indicator.
BATT
Low battery indicator.
Silenced
Alarm and alert tone silenced
indicator.
Keys
Nine control keys provided.
Brightness
Adjustable brightness of SpO
2
,
pulse rate, and bar graph dis-
plays.
Indicators
POWER
Indicates oximeter connected to
AC power.
CHGING
Indicates oximeter battery is
charging.
SpO
2
Range
0-100%
Accuracy
±2% at 70-100%
±3% at 50-69%
Alarm Limits
High 100-50% and off in 1%
steps.
Low 50-99% and off in 1% steps.
Averaging
4, 8, or 16 pulse beat average.
Sensors
Red
Infrared
660nm, 2mW (typical)
905nm, 2-2.4mW (typical)