PES:
PORTABLE ELECTROANESTHESIA SYSTEM
10
www.smith-root.com
TIMER, and VOLTS buttons, as previously
described for a single phase output.
Releasing the remote switch at any time
will stop the output of the PES.
In Dual-Mode the prefix “Pri” is displayed
for the Primary dose setting. Pressing
“SELECT” advances to display the prefix
“Sec” for the secondary dose. Pressing
“SELECT” again sets the displayed option
for the Secondary dose type.
Dual mode requires the pulse TIMER
for the primary waveform. Disabling the
primary waveform timer disables the dual
mode.
DEFAULT SETTINGS
The PES
®
comes with 10 sets of
parameters pre-programmed as default
doses. The individual parameters of the
doses can be changed. The 10 storage
locations (identified on the screen) can be
reviewed, changed, and stored in memory
for future use, by the user.
To restore the factory default settings
press the “MENU” button, press the “UP”
arrow to cycle to “System Menu 9/10”,
“Restore factory default settings?” and
then press “SELECT”. Press the “UP” arrow
to confirm the operation.
When powered on, the unit initializes with
the dose settings previously loaded by the
user. The dose settings to be used should
be reviewed prior to operation of the unit.
CONTROLS & SETTINGS