10600266 Rev. E

DYONICS™ Handpieces     

Instructions for Use

Cleaning and Sterilizing

Cleaning

CAUTION: Dried blood, saline, and other deposits 

inside the handpieces are a major cause of equipment 

malfunction.
CAUTION: Do not use an ultrasonic cleaner or any other 

automated sterilizing or cleaning equipment. 
1.  Dispose of the blade(s) used during the operation 

following standard procedures for disposal of 

biohazardous waste. 

 

WARNING: Only Smith & Nephew DYONICS™ 

Disposable Endoscopic Blades can be used with 

the DYONICS POWER Shaver System or the DYONICS 

POWER II Control System . The blades are intended for 

single use only . Do not resterilize . Do not lubricate 

blades . Discard all blades after use .

2.  Disconnect the handpiece cable from the front panel 

by pulling back the locking collar on the connector, 

then unplugging it from the control unit. Do not 

disconnect the cable from the handpiece.

3.  Place the protective cap on the connector end  

of the cable. 

4.  Set the suction control lever to the fully open position.
5.  Rinse thoroughly with warm tap water for a minimum 

of two minutes, making sure to irrigate all features of 

the device.

6.  Immerse the handpiece and soak for a minimum of 

five minutes using a neutral pH enzymatic detergent.

7.  After immersion of the device into the enzymatic 

cleaner, agitate the device to ensure all of the air is 

removed from the device lumen.

8.  While the unit is immersed, clean using the specified 

brushes from Brushes Corp. (or equivalent) in Table 3.
–  With the suction control lever in the open position, 

insert the lumen brush in the device spigot 

(proximal end of the device) and clean the lumen 

using twisting and in/out movements of the brush 

for a minimum of 30 seconds.

–  Reinsert the lumen brush from the distal end of 

the device end and clean the lumen from the 

opposite direction using twisting and in/out 

movements of the brush for a minimum of  

30 seconds.

–  Using the blade port brush clean the nose cavity 

using twisting and in/out movements for a 

minimum of 30 seconds.

–  Using the fork area brush clean the drive fork area 

for a minimum of 30 seconds.

Table 3 . Brush Identification

Handpiece REF

Lumen  

Brush

Blade Port 

Brush

Fork Area  

Brush

7205354, 7205355, 7205971, 
7209891, 7210542, 7210586, 
72200616, 72200617, 
72200872, 72202546

Brush 1

Brush 2

Brush 3

7205357, 72201500, 72201503

Brush 4

Brush 5

Brush 3

Brush 1

Bristles: Nylon, 3.5" area length, 0.35" diameter (no taper) 
Brushes Corp. PN 86-0516-1212, or equivalent*

Brush 2 

Bristles: Nylon, 2.25" area length, 0.75" diameter (tapered to 0.60") 
Brushes Corp. PN 86-1234-0006, or equivalent*

Brush 3 

Bristles: Nylon, 0.75" area length, 0.5" diameter (no taper) 
Brushes Corp. PN 99-9986-9999, or equivalent*

Brush 4 

Bristles: Nylon, 1.25" area length, 0.28" diameter (no taper) 
Brushes Corp. PN 86-00004-0009, or equivalent*

Brush 5 

Bristles: Nylon, 1.55" area length, 0.5" diameter (tapered to 0.35") 
Brushes Corp. PN 86-0212-1234, or equivalent*

* Equivalent must include bristle material, bristle area length, and   
  bristle diameter as specified.

9.  Scrub crevices and around hinged/mating surfaces 

with the blade port brush.

10. Rinse thoroughly with warm tap water for a minimum 

of 30 seconds, making sure to irrigate all features 

of the device. Ensure that the suction control lever 

is opened and closed repeatedly while rinsing the 

device lumen.

Cleaning and Sterilizing

Summary of Contents for 7205354

Page 1: ...DYONICS Handpieces Instructions for Use 10600266 Rev E ...

Page 2: ......

Page 3: ...r on the order of a physician EU Not for general waste Humidity Forward clockwise direction of rotation distal view Reverse counterclockwise direction of rotation distal view Oscillate alternating forward reverse direction of rotation Suction On fluid flow through handpiece unrestricted Suction Off fluid flow through handpiece blocked ...

Page 4: ...indications 5 Warnings 6 Precautions 6 Instructions for Use 8 Suction 8 Hand Control Handpieces 9 Non Hand Control Handpieces 9 Installing a Blade in a Handpiece 11 Installing a Blade in the Mini Motor Drive 11 Blade Window Lock 11 Forward Reverse Speed Shift 11 Cleaning and Sterilizing 12 Cleaning 12 Cleaning Verification 12 Storage 12 Sterilization Parameters 12 Service 13 Handpiece Cable Replac...

Page 5: ...blades for resection of soft and osseous tissue including but not limited to use in large articular cavities small articular cavities and for Functional Endoscopic Sinus Surgery FESS The FESS application is limited to small blades that are appropriate for the procedure Contraindications Smith Nephew DYONICS Handpieces and Blades should not be used with patients exhibiting ankylosis and also lackin...

Page 6: ...avoid patient burns do not place the handpiece on the patient when it is not in use Precautions U S Federal law restricts these devices to sale by or on the order of a physician Prior to each use inspect the device to ensure it is functioning properly and not damaged Do not use a damaged device In all cases the judgment of the physician in the best interest of the patient may take precedence over ...

Page 7: ...cted by one or more of the following measures Reorient or relocate this equipment the other equipment or both Increase the separation between the pieces of equipment Connect the pieces of equipment into different outlets or circuits Consult a biomedical engineer Environmental Protection This equipment contains electronic printed circuit assemblies At the end of the useful life of the equipment it ...

Page 8: ... for replacement The DYONICS POWERMINI Handpiece is not compatible with the DYONICS POWER Shaver System Suction is connected to a portal on the proximal end of the handpiece and removes fluid and debris drawn through the blade window In addition to removing fluids the suction cools and lubricates the motor drive Cutting takes place when the blade edge rotates across the blade s outer window The bl...

Page 9: ... 7205357 REF 7205971 and REF 72201503 are hand held motor drives that are electrically connected to the control unit via a 10 foot 3 meter cable The non hand control handpieces are activated by a footswitch These handpieces run a variety of DYONICS Disposable Endoscopic Blades The handpieces are designed to sense the blade type installed and communicate that information to the control unit which s...

Page 10: ...PM Range Default Range Default Range Default Range Default DYONICS POWER REF 7205354 7205355 Long Blades Straight and Curved TURBOWHISKER 100 3000 2000 100 3000 2000 500 3000 2000 1 9 1 Rev 1 8 2 Rev 8 2 Rev DYONICS POWERMAX REF 7210542 7210586 7209891 3 5 mm Curved Blade 100 3000 2000 100 3000 2000 500 3000 2000 1 9 1 Rev 1 8 2 Rev 8 2 Rev DYONICS POWERMAX ELITE REF 72200616 72200617 Curved HELIC...

Page 11: ...sition of the inner blade relative to the opening at the top of the outer blade Hand Control Operation Press and hold the Oscillate button on the hand control handpieces for one second Footswitch Operation Press and hold the Window Lock button on the footswitch to rotate the blade When the blade is in the best stopping position for the procedure release the Window Lock button Control Unit Operatio...

Page 12: ...ush in the device spigot proximal end of the device and clean the lumen using twisting and in out movements of the brush for a minimum of 30 seconds Reinsert the lumen brush from the distal end of the device end and clean the lumen from the opposite direction using twisting and in out movements of the brush for a minimum of 30 seconds Using the blade port brush clean the nose cavity using twisting...

Page 13: ...uct returned that has been serviced by an unauthorized third party repair shop and or sterilized with a sterilization method other than one approved by Smith Nephew and or damaged by contact with other mechanical devices such as shavers may incur additional costs regardless of warranty status It is not necessary to include accessory items i e power cords etc when returning a device for service The...

Page 14: ...s Forward Reverse Oscillate Blade Window Lock and Speed Shift Equipped with 10 foot 3 meter autoclavable power cord not field replaceable Suction control lever DYONICS POWERMAX and POWERMAX ELITE Handpieces REF 7210542 REF 72200616 REF 7220617 Length 7 1 18 0 cm Weight 7 5 oz 223 grams Lightweight handpiece with push button motor controls Forward Reverse Oscillate Blade Window Lock and Speed Shift...

Page 15: ...eight 5 0 oz 142 grams Standard handpiece with push button motor controls Forward Reverse Oscillate Blade Window Lock Equipped with 10 foot 3 meter autoclavable power cord not field replaceable Suction control lever Allows for shaver blades to be inserted with the blade window aligned to the suction control lever or 180 opposite the lever Specifications subject to change without notice Handpiece d...

Page 16: ... For a Return Authorization RA number or for additional information on this program call Customer Service at 1 800 343 5717 in the U S or contact your authorized representative 24 hour shipment is not offered in all countries Repair Service Program For devices no longer under warranty repairs can be made by Smith Nephew or by an authorized agent Non warranty repairs will be made at the list price ...

Reviews: