PSE57

 

Series

Pressure Sensor for General Fluids

How to Order

PSE57 0 01

Port size

Sensor range

Output specification

Specifications

Options/Part Nos.

Model

PSE570

PSE573

PSE574

PSE575

PSE576

PSE577

Fluid

Applicable fluid

Gas or liquid that will not corrode the materials of parts in contact with fluid

Pressure

Rated pressure range

0 to 1 MPa

−100 to 100 kPa

0 to 500 kPa

0 to 2 MPa

0 to 5 MPa

0 to 10 MPa

Proof pressure

3.0 MPa

600 kPa

1.5 MPa

5.0 MPa

12.5 MPa

30 MPa

Electrical

Power supply voltage

12 to 24 VDC 

±

10% with 10% voltage ripple or less

Current consumption

10 mA or less

Protection

Reverse connection protection

Accuracy

Analog output accuracy (Ambient temperature of 25

°

C)

±

1.0% F.S.

±

2.5% F.S.

Linearity

±

0.5% F.S.

Repeatability (Ambient temperature of 25

°

C)

±

0.2% F.S.

±

0.5% F.S.

Temperature characteristics 

(25

°

C reference)

±

2%F.S. (0 to 50

°

C)

±

3%F.S. (−10 to 60

°

C)

±

3% F.S. (0 to 50

°

C)

±

4% F.S. (−10 to 60

°

C)

±

5% F.S. (−10 to 60

°

C)

Environment

Enclosure

IP65

Withstand voltage

500 VAC for 1 minute between terminals and housing

Insulation resistance

100 M

Ω

 or more (500 VDC measured via megohmmeter) between terminals and housing

Operating temperature range

Operating: −10 to 60

°

C, Stored: −20 to 70

°

C (No freezing or condensation)

Operating humidity range

Operating/Stored: 35 to 85% RH (No condensation)

Standards

CE marking (EMC directive/RoHS directive)

Materials of parts

in contact with fluid

Piping port: C3604 + Nickel plating,

Pressure sensor: Al

2

O

3

 (Alumina 96%), O-ring: FKM + Grease

Piping port: C3604 + Nickel plating,

Pressure sensor: Al

2

O

3

 (Alumina 96%), Square ring: FKM

Model

PSE57

-

PSE57

-

-28

Analog 

output

Output

Voltage output: 1 to 5 V

Current output: 4 to 20 mA

Impedance

Output impedance: Approx. 1 k

Ω

Maximum load impedance: 500 

Ω

 or less (at 24 VDC)

100 

Ω

 or less (at 12 VDC)

Nil

Voltage output type 1 to 5 V

28

Current output type 4 to 20 mA

Symbol

Port size

Model

PSE570 PSE573 PSE574 PSE575 PSE576 PSE577

01

R1/8 (with M5 female thread)

02

R1/4 (with M5 female thread)

0

Positive pressure [0 to 1 MPa]

3

Compound pressure [−100 to 100 kPa]

4

Positive pressure [0 to 500 kPa]

5

Positive pressure [0 to 2 MPa]

6

Positive pressure [0 to 5 MPa]

7

Positive pressure [0 to 10 MPa]

Piping Specifications

Part no.

PSE570/573/574-01 PSE570/573/574-02

PSE575/576/577-02

Port size

R1/8

M5 x 0.8

R1/4

M5 x 0.8

R1/4

M5 x 0.8

Materials of parts

in contact with fluid

Piping port: C3604 + Nickel plating

Pressure sensor: Al

2

O

3

 (Alumina 96%)

O-ring: FKM + Grease

Piping port: C3604 + Nickel plating

Pressure sensor: Al

2

O

3

 (Alumina 96%)

Square ring: FKM

Weight

Without lead wire

and M12 connector

88 g

95 g

103 g

With lead wire

and M12 connector

175 g

182 g

191 g

Option (Lead wire)

Nil

Lead wire and M12 connector

(3 m), Straight

L

Lead wire and M12 connector

(3 m), Right angle

N

None

*

  See page 9 for connection to the PSE300AC.

Description

Part no.

Material

Note

q

Lead wire and M12 connector (3 m), Straight

ZS-37-A

1 pc.

w

Lead wire and M12 connector (3 m), Right angle ZS-37-B

1 pc.

e

Assembly-type connector

PCA-1557743

1 pc.

r

Adapter with restrictor Rc1/4

ZS-31-X175

Stainless steel 304

1 pc.

t

Adapter with restrictor Rc1/8

ZS-31-X188

1 pc.

y

Orifice M5

ZS-48-A Stainless steel 303

1 pc.

u

q

 + 

e

ZS-37-A-X448

The lead wire and connector 

are shipped together. (but not 

assembled)

i

w

 + 

e

ZS-37-B-X449

For pressure switch precautions and specific product precautions, refer 
to the “Operation Manual” on the SMC website. Click 

here

 for details.

3

C

Summary of Contents for PSE300AC

Page 1: ...e ring FKM Pressure Sensor 3 Screen Display Sensor Monitor Remote Type 0 to 1 MPa 0 to 2 MPa 100 to 100 kPa 0 to 5 MPa 0 to 500 kPa 0 to 10 MPa Measured value Current pressure value Label Display item...

Page 2: ...500 kPa PSE575 576 577 2 MPa 5 MPa 10 MPa For General Fluids Variations PSE56 Series Wetted parts Stainless steel 316L IP65 Oil free Single diaphragm construction Compact Pneumatic Pressure Sensor Lo...

Page 3: ...et value Threshold value Reverse output Set value Threshold value Normal output Hi side Set value Threshold value Hysteresis Set hysteresis value Reverse output Lo side Set value Threshold value Rever...

Page 4: ...Output Voltage output 1 to 5 V Current output 4 to 20 mA Impedance Output impedance Approx 1 k Maximum load impedance 500 or less at 24 VDC 100 or less at 12 VDC Nil Voltage output type 1 to 5 V 28 C...

Page 5: ...l 304 Lead wire and M12 connector ZS 37 A 1 to 5 VDC 4 to 20 mA DC mm Part no D E F G H I ZS 31 X188 20 9 R1 8 Rc1 8 14 1 5 ZS 31 X175 29 13 R1 4 Rc1 4 17 1 6 Internal Circuits and Wiring Examples PSE...

Page 6: ...voltage protection up to a voltage of 26 4 VDC Display Unit 3 MPa kPa Pa kgf cm2 bar mbar psi inHg mmHg mmH2O Display type LCD Number of screens 3 screen display Main screen Sub screen x 2 Display co...

Page 7: ...ng Examples Setting of NPN open collector 2 outputs Pressure sensor 3 wire type Setting of NPN open collector 2 outputs Pressure sensor 2 wire type Setting of PNP open collector 2 outputs Pressure sen...

Page 8: ...This eliminates wide variations of the indicated value Suction Verification Formula for Obtaining the Set Value This function constantly detects and updates the maximum minimum pressure when the powe...

Page 9: ...ple setting mode or F1 A large bore cylinder or ejector consumes a large volume of air during operation and may experience a temporary drop in the supply pressure This function prevents the detection...

Page 10: ...Straight 3 m ZS 37 B Right angle 3 m PCA 1557743 Assembly type connector ZS 31 C Right angle 5 m PLC PCA 1557743 Assembly type connector Lead wire and M12 connector Assembly type connector Set part no...

Page 11: ...ironments outside of the given specifications or use outdoors or in a place exposed to direct sunlight 2 Installation on equipment in conjunction with atomic energy railways air navigation space shipp...

Reviews: