Gas Range Adjustment Example
Enter
Range
S.P. Function - Range Adjust
*C
Enter
*100 PPM
Select [E] to select or
S
or
T
to select another and
press [E]
Down
10 PPM
Press [E] if selecting 0-10 PPM range
Down
USER
Press [E] if selecting User adjustable range
Enter
100 PPM
Select [E] to select or
S
or
T
to select another and
press [E]
4-20 mA Adjustment Example
Enter
Calib
S.P. Function - Calib Adjust
*D
Enter
Out: 4mA
Use
S
or
T
keys to select 4 mA or 20mA
Enter
4mA
Selects 4 mA
Enter
ACK
Momentary Acknowledge of new Set Point
Mode
Calib
S.P. Function - Calib Adjust
Down
CalibOut
S.P. Function - Output during Calibration Adjust
Enter
Track
Use
S
or
T
keys to select Track, Zero, C1.50mA, C4.00mA
Track = Output during calibration tracks the calibration gas,
Zero = Output is Zero during calibration,
C1.50mA = Output during calibration is 1.50 mA, C4.00mA =
Output during calibration is 4.0 mA
Enter
*Track
* = Current selection
RS-485 Adjustment Example
Enter
Address
S.P. Function - RS-485 Address Adjust
*E
Enter
Addr:016
Use
S
or
T
keys to enter new address
Enter
ACK
New address selected
Enter
Address
S.P. Function - RS-485 Address Adjust
Table 4-3B Set-Up Configuration
Enter
*38400
Press [E] to select or [
S
] or [
T
] to select another
Down
Baud
S.P. Function - RS-485 Baud Rate Adjust
Summary of Contents for 5100 IT Series
Page 2: ...Contents THIS PAGE INTENTIONALLY LEFT BLANK...
Page 10: ...Page 6 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 16: ...Page 12 Figure 3 4 4 20mA Circuit Types...
Page 17: ...Page 13 Figure 3 5 4 20mA Circuit Type Connections for 5100 XX IT...
Page 20: ...Page 16 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 27: ...Model 5100 XX IT Toxic Gas Sensor Module Page 23 THIS PAGE INTENTIONALLY LEFT BLANK...