3.3 Connecting to the Terminal

Remove the lid slowly and gently as there is a wire attached to it that serves as the screening 

connection for the lid.

Please also ensure that on replacing the lid that the wire does not get trapped in the sealing lip.

Connecting wire ends to the terminal is intended to be easy. Above each terminal block there is an opening 
and a metal clip inside. Use a small flat-bladed screwdriver (supplied) to press down firmly on the metal clip, 
and insert the exposed wire into the terminal access located on the side of the terminal block. Once satisfied  
that the wire is inserted deep enough, release the pressure, and the wire will be firmly secured in the 
terminal. Please see the figures below.

Using this system ensures that all wires have a constant pressure exerted on them,   maintaining a 

good connection. 

5

1 – 4 Channel External Amplifier

Take the wire that is to be inserted into 
the terminal entry in hand.

Press down firmly on the clip inside the 
opening on the top of the terminal 
block using a small flat-headed 
screwdriver.

Whilst pressing down, insert the bare 
wire end inhto the entry.

Remove the pressure from the clip to the 
bare wire. The wire should be firmly held 
in place. Test this by gently tugging the 
wire.

Summary of Contents for SKP 125 VS

Page 1: ...4 Channel External Amplifier SKP 125 VS Skye Instruments Ltd 21 Ddole Enterprise Park Llandrindod Wells Powys LD1 6DF UK Tel 44 0 1597 824811 skyemail skyeinstruments com www skyeinstruments com Iss 1...

Page 2: ...ems Feel free to contact us via our e mail or any of the methods below Click on the icons to browse to the sites or search for the usernames below Have a Smartphone Scan this QR code to access our web...

Page 3: ...ODUCTION 1 2 TYPICAL APPLICATION 2 2 1 Amplifier 2 3 WIRING POWER DETAILS 3 3 1 Wiring 3 3 2 Sensor Output Power Supply 4 3 3 Connecting to Terminal 5 4 HOUSING 6 5 SPECIFICATIONS 7 6 TROUBLESHOOTING...

Page 4: ...light can be accurately measured e g this may be useful when the maximum light levels are unknown at the outset Our range of amplified sensors have the amplification is built into the sensor housing w...

Page 5: ...X depending on the application It is possible to have different channels providing different output ranges for a given sensor but only one range would be available at any one time There is a maximum o...

Page 6: ...utput cable Grey 9 Sensor cable screens Grey 10 Sensor input grounds Green x 4 11 Channel 1 light sensor Input Blue Serial 12 Channel 2 light sensor Input Blue Serial 13 Channel 3 light sensor Input B...

Page 7: ...0 mol m 2 s 1 SKL 2633 Lux 0 1V 0 150 000 Lux 100 000 Lux SKL 2640 PAR Energy 0 1V 0 600 watts m 2 500 watts m 2 SKL 2650 Pyranometer 0 1V 0 1 500 watts m 2 1 100 watts m 2 If the sensor s is are comp...

Page 8: ...cated on the side of the terminal block Once satisfied that the wire is inserted deep enough release the pressure and the wire will be firmly secured in the terminal Please see the figures below Using...

Page 9: ...isconnected and care should be taken whilst removing the cover that the wire is not inadvertently tugged out of its place Dimensions of the amplifier unit are approx 85mm x 81mm x 36mm LxWxH Cables en...

Page 10: ...0 100 Temperature 20o C to 70o C Power Requirements 0 up to 3V output range 5 15V power supply 0 3 6V output range 9 15V Outputs Custom range specified at time of ordering e g 0 1V 0 5V etc Scaling Cu...

Page 11: ...rminal that has been configured for them Also check that the four output wires are connected to the correct channels in the measuring equipment and that the system is interpreting the output in the co...

Page 12: ...Skye Instruments Ltd 21 Ddole Enterprise Park Llandrindod Wells Powys LD1 6DF UK Tel 44 0 1597 824811 skyemail skyeinstruments com www skyeinstruments com...

Reviews: