PRODUCT SPECIFICATIONS
SKF Multilog On-line System IMx-8/IMx-8Plus
46 (63)
SKF Multilog On-line System IMx-8/IMx-8Plus
User Manual
Revision E
3.2 SKF Multilog On-line System IMx-8/IMx-8Plus
3.2.1 Environmental
•
DIN rail mounting (35 mm x 7.
5 mm ‘top hat’ DIN
rail)
•
Size (H x W x D): 172 x 104 x 40 mm (6.8 x 4.1 x 1.6 in.)
o
Height (H) across, width (W) along and depth (D) above the rail
o
Height does not include terminal connectors and Bluetooth dongle
o
Depth does not include the DIN rail mounting bracket
•
Device weights: IMx-8: 465 g (1.03 lb) and IMx-8Plus: 580 g (1.28 lb)
•
IP ratings:
o
IP30, for the DIN rail mounted enclosure
o
IP65, for the wall mounted SKF cabinets
o
IP67, IP69K and IK09 for the optional mobile data/Wi-Fi antenna
o
IP69K for the optional GPS module
•
Temperature ranges
o
Operating:
IMx-
8: −40 to +70 °C
(−40 to +158 °F)
IMx-8Plus:
−
40 to +65 °C
(−40 to +149
°F)
o
Storage:
−
50 to +85 °C
(−
58 to +185 °F)
•
Humidity: 95% (relative) non-condensing
•
Measurement category II
•
Pollution degree 2
•
Maximum altitude: 2 000 m (6 562 feet)
•
Vibration tolerance:
4
–
13.2 Hz 1 mm
13.2
–
100 Hz 0.7 g
Number of axes: 3 mutually perpendicular
3.2.2 Power
•
10 W or less typical, 13 W maximum
o
Power over Ethernet (PoE) max 13 W (nominal voltage 48 V) or
o
24 to 48 V DC nominal (range 22 to 50 V DC)
o
Supply fuse rating: T2AL (2 A slow blow)
o
PoE and 24 to 48 V DC can be used redundantly
•
Optional CMON 4137: DIN rail mounted power supply, suitable for all
IMx-8/IMx-16
devices including ‘Plus’ versions
o
Universal AC input: 100 to 240 V AC (nominal)
o
DC input range: 120 to 370 V DC
o
Dimensions: 74 x 54 mm (2.91 x 2.13 in.) across/along the rail
3.2.3 Analogue inputs
•
8 analogue single ended inputs referenced to chassis/enclosure ground
o
Input range ±25 V
o
Supports 2-wire constant current sensors and/or voltage signals
o
4-20 mA signals accepted (required load resistor to be fitted at input)
•
Software controlled power supply for standard accelerometers