Functional description and operating instructions
Settings
2 - 4
Wireless Infrared Thermometer CMAC 4230
User Manual
Settings
Open the battery compartment. You can set the DIL switches.
Lock:
(ON) locks the measurement button.
°C / °F:
Toggling the display between °C and °F (the
time and date format also change).
Buzzer:
(ON) acoustic alarm.
Backlight:
(ON) display light
Set default:
Reset to the factory defaults.
Ltd. Access:
Function keys are disabled.
Laserflash:
The laser marking starts blinking when the
alarm values are exceeded.
Auto Off:
(On) The turn-off time does not extend
automatically after receiving a command from
the software (applicable only to the operation in
the master mode)
Slv./Mstr.:
Toggles the BluetoothTM mode.
Time/Date:
Toggles between the date and time display.
TC-K:
Thermocouple type
Temperature range for the K type: -30 to +400
°C
The thermocouple must be used only outside the Hazardous area!