Mounting Sampling Media
Sampling with a Filter Cassette or
BioStage 200 Impactor
Use the tubing adapter supplied with the pump
(see right)
.
Insert the smooth end of the tubing adapter into the pump
inlet up to the flange. Attach 3/8-inch ID tubing to the
barbed section of the adapter. Attach tubing to the outlet
of a sampling cassette, cassette holder, or BioStage 200
Impactor.
Sampling with Spore Traps
Use the Cassette Adapter supplied with the
pump
(see right)
.
Insert the tube end of the adapter into the
pump inlet up to the flange. Remove the seal
from the outlet of a spore trap cassette and
push the cassette onto the bowl end of the
adapter until a firm seal is established.
Calibration Note:
Flow rate can be calibrated
by pushing the bowl end of a second cassette
adapter onto a spore trap cassette inlet (seal
removed), attaching flexible tubing to the
tube end of the second adapter, and attaching
the other end of the flexible tubing to the
outlet of a calibrator.
Sampling with Sampling Wand Accessory (Cat. No. 228-9521)
Insert the tube end of a cassette adapter into the hole on the top end of the sampling
wand accessory up to the flange (
see above
). Push the outlet of a spore trap cassette
onto the bowl end of the adapter until a firm seal is established. Insert the tubing
adapter on the flexible tubing (supplied with wand) into the pump inlet up to the
flange.
Spore trap
cassette
Cassette
adapter
Wand
Tubing
included
with wand
Tubing
adapter
included
with wand
Empty cassette
adapter
Cassette adapter
with spore trap
cassette