Attaching/Removing the Preamplifier
To attach the preamplifier (
see below
), position the RED orientation key on the preamplifier with the RED
identification mark on the sound level meter and gently push the preamplifier into the connector.
To remove, gently pull the preamplifier stem away from the instrument body.
Certifiable Calibration
The calibration process includes the microphone, preamplifier and sound level meter. Any change in this
measurement chain will require a new calibration certificate.
SKC offers a complete calibration service offering a NIST calibration that can be supplied with or without
a test report.
It is recommended that your sound meter instrumentation is calibrated annually to ensure your measuring
equipment is completely accurate and fully compliant
.
Measuring Sound
•
Always calibrate your SoundCHEK Essential sound level meter prior to and after taking
measurements using an SKC AcoustiCHEK Calibrator Cat. No. 703 Series
.
Sound Level
– General Recommendations
In some environments, high levels of noise can occur. The SoundCHEK Essential sound level meter has
been designed for complete accuracy up to sound levels of 140 dB.
Before you record measurements, take the time to ensure you have chosen the optimum range for the
application. Wherever possible, the optimum range is when the average measured signal is
approximately halfway between the top of the range and the bottom of the range without an overload
condition.
Preamplifier
Microphone
Only pull out and plug
in the preamplifier.
DO NOT TWIST