Language
Select the desired operating language of the instrument by scrolling through the options using the
Up
or
Down Arrow
keys and pressing
OK
to confirm the selection.
Calibration
Select Calibration to acoustically calibrate your SoundCHEK Essential sound level meter.
It is recommended that the calibration procedure is performed before and after measurements have been
taken using an SKC AcoustiCHEK Calibrator Cat. No. 703 Series with an equivalent class rating (or
better) to match the model of your SoundCHEK Essential sound level meter.
Select
User Calibration
to allow access to the calibration routine and allow the user to set calibration
settings.
Calibrate to
1. Adjust the
Calibrate to
value between 91.0 and 120.0 dB in 0.1-dB steps using the
Left
or
Right
Arrow
keys or 1-dB steps using the
Up
or
Down
Arrow
keys. Only include the microphone-
dependent pressure to free field correction in this calibration adjustment.
Ensure the calibrator is attached to the microphone by gently inserting the microphone into the
cavity of the calibrator. A certain amount of resistance should be felt while inserting the
microphone as the O-ring on the calibrator forms a seal around the microphone. Switch the
calibrator on and set to the required output level.