Sampling
•
Allow pump to equilibrate after moving it from one temperature extreme to another.
•
Charge pump battery completely before sampling.
•
Use of an unapproved battery and/or charging cable could damage the pump and will void any
warranty.
•
Use of any device (including charging cradle) or battery pack other than Cat. No. P75718 to
power the pump voids intrinsic safety certifications and any warranty.
•
Pump can be operated from cradle.
•
If using sample tubes as media, calibrate/verify pump flow rate before and after each sampling
operation using the tube holder and pump used for sampling.
1. After setting/calibrating flow rate, ensure that calibrator and tubing have been removed and
representative method-specified sample medium used for calibration has been replaced with
newly opened unexposed method-specified sample medium to complete the sampling train.
See
Figure 8
.
2. Choose from a manual sample, timed sample, or sample preset (presets are uploaded to the
pump from DataTrac Pro Software).
See Manual Sample, Timed Sample, or Presets below.
Figure 8. High Flow Sample
Train
Sample medium
Tubing
Pump inlet