Recommended Separation Distances between portable and mobile RG Communications
equipment and bed.
Equipment and Systems that are NOT Life-Supporting
Recommended Separations Distances for the SW Evolution™ and Rest Secure System™: The
SW Evolution™ and SW Rest Secure System™ are intended for use in the electromagnetic
environment in which radiated disturbances are controlled. The customer, or user, of the SW
Evolution™ and SW Rest Secure System™ can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF Communications Equipment
and the bed, as recommended below, according to the maximum output power of the
communications equipment.
D=(3.5/V1)(Sqrt P)
D=(3.5/E1)(Sqrt P)
80 to 800 MHz
D=(7/E1)(Sqrt P)
800 MHz to 2.5 GHz
Compliance Level Cond RF 3 Rad RF-800 MHz 3 Rad RF – 2.5 GHz
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.69
3.69
7.38
100
11.67
11.67
23.33
General Safety
WARNING:
The bed is to be used in accordance with each facility’s policies
and procedures.
CAUTION:
Possible fire hazard when used with oxygen administering
equipment other than the nasal, mask or 1/2 bed length type tent. Oxygen tent
should not extend below the mattress support level. Lock hand control at foot of
bed when using oxygen administering equipment.
WARNING:
Electric shock may occur when plugging the bed into the wall
outlet. Use ONLY grounded or hospital grade outlets.
DO NOT use the bed if the power cord is cut, frayed or loosely connected to the
bed.