7 Conditioning
Operating instructions VDW 6:1
32
61 34 378 D3451
.
201.04.06
.
02
NOTICE! Never clean in an ultrasound bath!
NOTICE! Never immerse in disinfectant solution!
Using the T1 spray
NOTICE! Only use Sirona T1 spray.
Use soft brush
IMPORTANT: Use a soft, clean, and disinfected brush for cleaning.
Procedure _ Instrument without integrated spray
✔ Wear appropriate protective clothing.
✔ All disinfectants must be approved in your country and have proven
bactericidal, fungicidal and virucidal properties. Only use
disinfectants with no protein-fixing properties.
1. Brush the instrument under running water (< 38 °C, < 100 °F, at
least drinking water quality) and good lighting (min. 500 lux) and
color rendering index (min. 80 Ra) until no more dirt can be seen, for
at least 10 seconds.
2. Flush the drive channels with spray.
3. Conduct thermal disinfection or unwrapped steam sterilization.
4. Maintain mechanical parts manually [ → 33].