Operating instructions VDW 6:1
7 Conditioning
61 34 378 D3451
.
201.04.06.02
31
1. Check whether the instrument is clean after conditioning under good
lighting (min. 500 lux) and color rendering index (min. 80 Ra).
2. If it is still dirty, repeat the process.
For further conditioning, the instrument should be dry and free of
residue.
3. Blow the instrument out with max. 3 bar.
4. Manual care of mechanical parts [ → 33].
5. Manual care of the push button chuck [ → 35].
6. Pack the instrument in packaging material suitable for sterilization
and storage. e.g. paper/plastic composite packaging.
7. Sterilize the instrument [ → 36].
7.3
Manual cleaning and disinfection
Immediately after treatment
NOTICE! Condition immediately, or at the latest, one hour after
treatment.
6134378_EN.book Seite 31 Montag, 10. August 2015 12:17 12