Sirona Dental Systems GmbH
Operating Instructions SIROLaser Xtend
9 Electromagnetic compatibility
9.1 Electromagnetic emission
63 18 351 D3576
D3576.201.01.05.02 11.2018
69
9
Electromagnetic compatibility
NOTE
The SIROLaser Xtend complies with all requirements for
electromagnetic compatibility according to IEC 60601-1-2: 2007
Definitions:
Emission (electromagnetic)
When electromagnetic energy is emitted by a source.
Interference immunity
The ability of a device or system to work without errors even if there is
electromagnetic interference.
Immunity level
The maximum level of a certain electromagnetic interference that
affects a particular device or system, where the device or system
remains operative with a certain level of performance.
9.1
Electromagnetic emission
The UNIT is intended for operation in the electromagnetic environment
specified below.
The customer or user of the UNIT should make sure that it is used in
such an environment.
Emission measurement
Conformity
Electromagnetic environment - guidelines
RF emissions according to CISPR 11
Group 1
The UNIT uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
RF emissions according to CISPR 11
Class B
The UNIT is intended for use in all facilities, in-
cluding residential areas and in any facilities con-
nected directly to a public power supply providing
electricity to buildings used for residential pur-
poses.
Harmonics
according to IEC 61000-3-2
Class A
Voltage fluctuations / flicker according
to IEC 61000-3-3
coincides