The following is a typical example of a treatment submenu.
1. Selected program
in this example Frenectomy
2. Laser power
in our example, 3.0 W power. By pushing on the power touch field you
will be transferred to another screen where you will be able to adjust the
emitted power between 0.5 W and 5 W in 0.1 W increments either by
typing in the numbers or by moving higher or lower by using 'plus' or
'minus'. After the upgrade you will be able to adjust the emitted power
up to 7 W.
WARNING
The preset power levels are considered to be safe for patients.
Increasing the power levels entails the risk of overheating the
patient's soft or hard tissue. Setting the power to excessively low
levels may result in reduced treatment efficacy.
3. Time
in our example, continuous. The laser will be activated as long as the
finger switch is pressed. By pushing on the time touch field you will be
transferred to another screen where you will be able to adjust the time
between continuous or 1 to 9999 seconds either by typing in the
numbers or by moving higher or lower by using 'plus' or 'minus'.