60 78 104 D 3488
D 3488.201.01.07.02
11.2009
33
Operating Instructions SIROEndo
10 Electromagnetic compatibility
båÖäáëÜ
10
Electromagnetic compatibility
10.1
Electromagnetic emission
The
UNIT
is intended for operation in the electromagnetic environment spec-
ified below.
The customer or user of the
UNIT
should make sure that it is used in such an
environment.
Emission measurement
Conformity
Electromagnetic environment guidelines
RF emissions according to
CISPR 11
Group 1
The
UNIT
uses RF energy only for its internal function. There-
fore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF emissions according to
CISPR 11
Class B
The
UNIT
is intended for use in all facilities, including residen-
tial areas and in any facilities connected directly to a public
power supply providing electricity to buildings used for resi-
dential purposes.
Harmonics
according to
IEC 61000-3-2
not applicable
Power output
< 50 W
Voltage fluctuations/Flicker
according to
IEC 61000-3-3
not applicable
no significant
flicker