63 22 668 D3561
10
D3561
.
021.01.06
.
02 12.2014
3 On-site installation
Sirona Dental Systems GmbH
3.1 Substrate, floor
Installation Requirements SINIUS / SINIUS CS / SINIUS TS
3.1.3
Stability
With a concrete floor/SINIUS
With a concrete floor
● Use a 12 mm dia. masonry drill to drill the holes.
● The depth of the drill hole (C) in the concrete floor must be at least 80
mm.
● The specified drilling depth (C) of at least 80 mm refers to the drilling
length without screed (A) or impact sound insulation (B). The heavy
duty anchor bolts included with delivery (2 units) are suitable for a
clamping length ((A)+(B)) of up to 100 mm.
● Use 12 mm dia. wall plugs (but do not use wall plugs instead of heavy
duty anchor bolts).
● Use 10x160 mm wood screws including washers.
With a wooden floor
With a wooden floor
● Provide and prepare appropriate supporting beams.
● Use a 7mm dia. wood drill to drill the holes.
● The depth of the drill hole must be at least 80mm.
● Use 10x80mm wood screws including washers.
ATTENTION! Do not use wall plugs!
Ø 12mm
Ø 12mm
20mm
7mm
160
mm
160
mm
80
mm
min.
80
mm
min.
C
B
A
100
mm
max.
100
mm
max.
WARNING
Any existent screed (A) and/or impact sound insulation (B) must be
penetrated!
Ø 7mm
Ø 7mm
7mm
80
mm
min.
80
mm
min.
Summary of Contents for Sinius
Page 1: ...f o NOKOMNQ pfkfrp L pfkfrp p L pfkfrp qp k W b Installation requirements...
Page 2: ......
Page 57: ......