1.1 Technical Data
60 82 387 D 3285
D 3285.201.05.04.02
5
båÖäáëÜ
1
Technical Description
1.1
Technical Data
ORTHOPHOS 3 Ceph has been inspected by the
VDE Testing and Certification Institute for compli-
ance with EN 60601-1, EN 60601-2-28 and EN
60601-1-3 and has been found to comply with these
regulations
.
Original language German
This product is provided with a
CE
marking in
accordance with the regulations stated in the Directive
93/42/EEC of June 14, 1993 concerning medical
products.
Exposure times tele-exposure:
0,1 – 4,00s
Automatic Exposure Blockage:
The duration of the exposure blockage (cool-off period)
depends on the kV/mA step set and the actually triggered
radiation time. Depending on the tube load, pause times up
to 200s are set automatically.
Example: For tele-exposure with exposure data 78kV/10mA
and exposure time 0.8s a cool-down time of 8s is observed.
Source - Image receptor distance: Ceph
1660mm
Cassette format:
Ceph
USA/CANADA/
Corea
18cm x 24cm (7" x 10")
24cm x 30cm (10" x 12")
8" x 10"
10" x 8"
Diaphragm must fit to cassette.
0123
Reg. No.: China