4 Operation

Operating Instructions ORTHOPHOS 3 Ceph

4.2 Positioning the Patient for Tele-Exposures

60 82 387 D 3285
D 3285.201.05.04.02

13

båÖäáëÜ

Scale (26) now gives a reading.

Use adjustment wheel 

A

 to set same value on 

soft-tissue filter of tele-radiograph diaphragm No.3 
(only for asymmetric exposures).

Special setting

:

Adjust as required for attenuated or pronounced 
recording of soft tissue.
For attenuated recording: set upwards
For pronounced recording: set downwards.

Selecting the tele-exposure data.

Briefly press one of the four program keys.
The LED above the key illuminates. 

The corresponding programmed 

kV/mA

 values and 

exposure time 

s

 are displayed.

Exposure data can be manually changed as follows:
Press key + or – until the desired 

kV

 or 

s

 value is 

displayed.

The LED over the patient symbol key does not light up in 
this case.

The patient symbol keys are preprogrammed in the 
factory with 

kV/mA

 and 

s

 values.

„Program Values – Tele-Radiography” on page 17. 

If you want to change them, see chapter
„Programming the Tele-Exposure” on page 16

26

120

110

100

90

80

70

60

mm

        

120

110

100

90

80

70

60

3

A

c

s

I0

74

0.64

c

s

manuell

Summary of Contents for orthophos 3 ceph

Page 1: ...b loqelmelp P l f q J b...

Page 2: ...ease take the ORTHOPHOS 3 Operating Instructions into account also as well as the Warnings and Safety Notes included therein Ihr ORTHOPHOS Team Maintenance To ensure the safety of the patient the oper...

Page 3: ...r 8 3 Accessories 9 3 1 Accessories for Cephalometer 9 3 2 Adjusting Film 9 4 Operation 10 4 1 Preparing the Tele Exposure 10 4 2 Positioning the Patient for Tele Exposures 12 4 3 Releasing the Tele E...

Page 4: ......

Page 5: ...th the regulations stated in the Directive 93 42 EEC of June 14 1993 concerning medical products Exposure times tele exposure 0 1 4 00s Automatic Exposure Blockage The duration of the exposure blockag...

Page 6: ...witch 2 Patient positioning mirror 3 Head holder with adjusting knob for temple supports 4 Cassette holder with carriage 5 Diaphragm 6 Diaphragm wheel with locking button 7 Height adjustment buttons A...

Page 7: ...splay for exposure program exposure time with buttons for exposure programs Exposure time for tele radiography Digital display for kV mA paired values with keys for overriding kV values Patient symbol...

Page 8: ...25 Projection scale 26 Nose support position scale horizontal 27 Ear olives with holders 28 Film cassette for tele radiography Reproducibility of Tele Exposures After positioning the patient note the...

Page 9: ...W may be disinfected with petroleum ether We recommend that you attach hygienic protective covers before each exposure 3 2 Adjusting Film The supplied adjusting film is needed for the maintenance and...

Page 10: ...mmetrical S for a p p a and carpus exposures Adjust diaphragm by pressing locking button and turning diaphragm wheel up to stop The diaphragm number appears in the upper right corner of the diaphragm...

Page 11: ...top and push out to side The cephalometer can be rotated The rotational apparatus locks in the 0 45 turning to the right and 90 position 0 position is for lateral exposures For adjustment the swivel l...

Page 12: ...or sit under cephalometer Press buttons or on unit to adjust cephalometer until both ear olives are at same height as external auditory canals Introduce ear olives into external auditory canals Hold...

Page 13: ...rding set downwards Selecting the tele exposure data Briefly press one of the four program keys The LED above the key illuminates The corresponding programmed kV mA values and exposure time s are disp...

Page 14: ...5 04 02 Slide film cassette into cassette holder up to lateral cassette stop locks facing cassette holder not pati ent Front cassette stop A asymmetric for lateral projections Back cassette stop S sym...

Page 15: ...ED goes out If no further exposures are to be made switch unit off Interrupting an Exposure If the exposure key is released prematurely exposure is interrupted The kV mA displays and the LED above key...

Page 16: ...on page 17 1 Adjust diaphragm 3 or 4 for ceph function The C dis play above s on the Multitimer illuminates 1 Set desired kV and s values with keys 2 Press memory programming key The LED above the pr...

Page 17: ...e developing equipment and instructions on the chemicals 6 1 Program Values Tele Radiography world wide except USA and Asia Intensifying Screen Kodak Lanex Regular with film Kodak T Mat G or Agfa Orth...

Page 18: ...m film and development tolerances may be compensated by changing the exposure data in one step up or down Table 5 Index 2A Factory programmed values with a film density of 03 Freely programmed values...

Page 19: ...om film and development tolerances may be compensated by changing the exposure data in one step up or down Table 8 Index 1A Factory programmed values with a film density of 03 Freely programmed values...

Page 20: ...t K p a p d e OMMN p W m d a POURKOMNKMRKMQKMO MPKOMMT KJk KW NMU ROP f p a p d e rp W W c PN SQSOR _ d K K p a p ii QUPR p a I p NMM I k OUOTP rp p PORM o a J r R j I l iRi RvS l k SM UO PUT a POUR...

Reviews: