Default settings
Operating Instructions inFire HTC
Adjustment of the start time (casting time function)
båÖäáëÜ
7.2
Adjustment of the start time (casting time
function)
Explanation
For calculation of the start time (when using the casting time function), the
furnace controller assumes an average voltage of 230 V.
The local voltage may significantly differ from this value according to local
circumstances.
Accordingly, the calculation of the start and end time of the controller may be
incorrect.
Substantial variations (of more than 30 minutes) may be adjusted by means
of the correction program.
Correction program
The correction program compares the values in a test run and automatically
corrects them.
This correction also remains intact even after the furnace is switched off.
Activation
1. Load program
00
.
2. Start the program.
ª
The program automatically switches off following the correction
measurement.
CAUTION
Furnace empty?
No sintering trays may be located in the furnace during the correction
program run.
NOTICE
Long program run
This program takes about 3-4 hours to run.
➢ You can quit the program by pressing the START/STOP key if
necessary.