FOR Safe AQ Voice meter
:
1. If test results greater than 600 mg/dL
(
33.3mmol/L
)
, the meter will display
“HI” and read“The result is beyond the range”with red light on. Retest your blood
glucose immediately with a new test strip. If your reading is still “HI”, contact
healthcare professional immediately.
2. If test results is lower than 20 mg/dL
(
1.1mmol/L
)
, the meter will display “LO”
and read “The result is below the range”with red light on. Retest your blood
glucose immediately with a new test strip. If your reading is still “LO”, contact
healthcare professional immediately.
3. If test result is >70mg/dL
(
3.9 mmol/L
)
and
<
126 mg/dL
(<
7.0 mmol/L
)
,
the meter will display and read the test result with green light on.
4. If test result is ≥ 126 mg/dL
(
≥7.0 mmol/L
)
and ≤ 250 mg/dL
(
≤13.9 mmol/L
)
,
the meter will display and read the test result with yellow light on.
5. If test result is ≤ 70mg/dL
(
≤ 3.9 mmol/L
)
or >250mg/dL
(
> 13.9 mmol/L
)
,
the meter will display and read the test result with red light on.
Note: Do not change your treatment only based on test result and light indications.
Please consult healthcare professional for your treatment.
V.About Your Safe AQ Smart
、
Safe AQ Air
、
Safe AQ Voice Meter
Test
strip
port
Strip
eject
b
utton
LCD
display
Main
button
Test
strip
port
Strip
eject
b
utton
LCD
display
Main
button
Battery Cover
Battery Cover
Safe AQ Smart Meter:
Safe AQ Air Meter:
Summary of Contents for Safe AQ Smart
Page 26: ...45...