Test Principle
A blood glucose test is based on measurement of electrical current caused by the
reaction of glucose with the reagents (special chemicals) on the electrode of the
strip. The blood or control solution sample is drawn into the tip of the test strip
through capillary action. Glucose in the sample reacts with the special chemicals
and generates electrons, which produce electrical current. The blood glucose
meter measures the electrical current and calculates the glucose result. The
glucose result is displayed by your meter as mg/dL or mmol/L.
II.Important Safety Information
▶
For self-testing customer, your Safe AQ Smart
、
Safe AQ Air
、
Safe AQ Voice
meter and lancing device are for single person use only. Do not share them with
anyone else, even family members.
▶
For professional use, if the meter is being used on the second person, the meter
and lancing device should be disinfected first.
▶
Safe AQ test strip and lancet are for single use only. DO NOT REUSE.
▶
Do not use other test strips and control solutions with the Safe AQ Smart
、
Safe
AQ Air
、
Safe AQ Voice meter.
▶
Do not use the meter if it is not working properly, or there is any damage.
▶
Keep the Safe AQ test strip vial away from children. The strip vial, test strips,
and control solution bottles can be choking hazards. DO NOT drink the control
solution.
▶
Remove the batteries if the meter is not likely to be used for some time.
DO NOT CHANGE YOUR TREATMENT BASED ON A SINGLE RESULT THAT
DOES NOT MATCH HOW YOU FEEL OR IF YOU BELIEVE THAT YOUR TEST
RESULT COULD BE FALSE.