Questionable Results:
X.Blood Glucose Testing
Prepare meter, strip, lancing device and lancet before test.
Probable Cause
Operation mistake
Control solution not being shaken well
Use the first drop of control solution
Expired or contaminated control solution
Expired or contaminated test strip
Temperature is too high or too low
Retest according to correct steps
Shake control solution well and retest with
a new test strip.
Retest with a new test strip
Change a new vial control solution and retest.
Change a new vial test strip and retest.
Equilibrate meter, test strip and control
solution in testing condition for 30 minutes
and retest.
Meter malfunction
Contact customer service or local dealers
Action