Simpson 385-3L Operator'S Manual Download Page 10

10

5.

SERVICING INSTRUCTIONS

This section will describe the necessary procedures needed to effectively
service this Instrument.

5.1

Battery Replacement

Replace the battery only if unable to adjust reading to 32

o

F as in paragraph

4.3, item  3. To replace the standard “C” size flashlight battery, remove the four
screws on the bottom of the case which holds the lower panel in place. Slide
the lower panel away from the Instrument cover about 1/32 inch to release
stop on top edge of panel, then lift the panel free of the case to gain access to
the battery. (Use insulated type batteries only.)

5.2

Calibration

If the Instrument is known to require calibration, or if the accuracy of the cali-
bration should be checked, proceed as follows:
1.

With the function switch OFF, check for a reading of -50

o

F. If necessary,

correct reading with zero adjusting screw on the face of the Instrument.

2.

Set the function switch to the READ position.

3.

Using an ice point reference source take a temperature reading.

4.

If the reading is not exactly 32

o

F, turn the slotted adjusting control below

the Simpson logo until the meter indicates 32

o

F, exactly.

5.

Turn the function switch to the ADJ position. The meter should still indi-
cate 32

o

F.

5.3

Care

1.

Immediately clean all spilled materials from the Instrument and wipe dry.
If necessary, moisten a cloth with soap and water to clean plastic sur-
faces.

2.

Whenever possible, avoid exposure or usage in areas which are subject
to temperature and humidity extremes, vibration or mechanical shock,
dust or corrosive fumes or strong electrical or electromagnetic interfer-
ences.

3.

Monthly Care: Verify Instrument calibration by performing operational
checks using known value sources. If the need for calibration is indicated,
contact the nearest Authorized Service Center.

4.

Annual Care: It is recommended that the Instrument be returned annu-
ally for a complete overall check and calibration.

5.

Storage: When the Instrument is not in use, store it in a location free from
temperature extremes, dust and corrosive fumes, and mechanical vibra-
tion or shock.

5.4

Replacement Parts

Table 5-1 lists parts in order of their reference designators and indicates their
description. (Refer to Table 1-2 for Items and Accessories Furnished With This
Instrument.)

Summary of Contents for 385-3L

Page 1: ...Model 385 3L V2 Temperature Tester OPERATOR S MANUAL...

Page 2: ...hnical Support or Customer Service at 847 697 2260 Internet http www simpsonelectric com Warranty and Returns SIMPSON ELECTRIC COMPANY warrants each instrument and other articles manufactured by it to...

Page 3: ...5 NOTE The proper measuring instrument for the measurement of leakage current consists essentially of a network of a 1500 ohm non inductive resistor shunted by a 0 15 microfarad capacitor connected be...

Page 4: ...ction 6 2 2 Warranty 6 2 3 Shipping 6 2 4 Operation Instructions 6 3 CONTROLS CONNECTORS INDICATORS 7 3 1 Front Panel 7 4 OPERATION 8 4 1 Safety Precautions 8 4 2 Test Probes 8 4 3 Preparation For Use...

Page 5: ...s information is flagged by CAUTION or WARNING symbols throughout this publication and is defined on the inside front cover of this manual under SAFETY SYMBOLS Adhere to these instructions in order to...

Page 6: ...ompany warranty policy is printed on the inside front cover of this manual Read it carefully before requesting a warranty repair For all assistance including help with the Instrument under warranty co...

Page 7: ...anel Front Table 3 1 Controls Connectors Indicators 1 Zero Adjusting Screw If pointer is off zero adjust screw slowly clock wise or counter clockwise until the pointer is exactly over the black diamon...

Page 8: ...sistor may result in a capacitor retaining a dangerous charge Turn off power and discharge all capacitors before connecting or disconnecting test leads to and from the circuit being measured 4 Hands s...

Page 9: ...sired 2 When measuring air temperature allow several minutes for the tempera ture of the probe sensing element to stabilize before taking a reading Liquid temperatures can be read within a few seconds...

Page 10: ...urn the slotted adjusting control below the Simpson logo until the meter indicates 32o F exactly 5 Turn the function switch to the ADJ position The meter should still indi cate 32o F 5 3 Care 1 Immedi...

Page 11: ...1k 5 5 119351 R6 Resistor 3 31k 0 5 6 114951 R7 Potentiometer 50k 20 6 114974 S1 Switch Probe Selector 6 114965 S2 Switch Function Selector 6 114966 M1 Meter 0 50 DC A 10 865868 Knob Function Probe Se...

Page 12: ...12 SIMPSON ELECTRIC COMPANY 853 Dundee Avenue Elgin IL 60120 3090 847 697 2260 FAX 847 697 2272 Printed in U S A Part No 06 114985 Edition 4 10 02 Visit us on the web at www simpsonelectric com...

Reviews: