Appendix B.
LCD Touch Screen Position Calibration
The SL-881 AVG LCD touch screen has been calibrated before shipping out. In case you need to calibrate touch screen
position, follow steps below. All the operation in this section is performed on the SL-881.
1.
Press the UPG button (item 11 in
) first, and then press the Reset button (item 13 in
) to reset
the SL-881. After the SL-881 completes resetting, the Touch Screen Calibration window appears. See
Touch Screen Calibration
At first time, you need to
calibrate the Touch Screen.
Please press the touch screen to
continue...
Figure B.1. Touch Screen Calibration Window Shown on the SL-881 LCD Touch Screen
2.
Touch any position on the touch screen to start the calibration.
3.
A message appears on the top left corner on the touch screen (
). Follow the instruction, press the circle.
Runtime calibration,
please touch the screen
at the center of the ring.
Press here
(upper left position)
Figure B.2. Click Circle on the Top Left Corner
4.
After that, another message appears on the lower right corner on the touch screen (
). Follow the
instruction, press the circle. The touch screen position calibration is finished.
Runtime calibration,
please touch the screen
at the center of the ring.
Press here
(lower right position)
Figure B.3. Click Circle on the Lower Right Corner
5.
After the position calibration is completed, the SL-881 will go back to the LCD Touch Screen main window
(
).