Register
Write
Range
Register description
A5h
1
Yes
0 ÷ 3
“
AL
” parameter in “
OUtP
” menu (current output value on critical
exception):
0
- no change;
1
- 22.1 mA;
2
- 3.4 mA;
3
- 0 mA
B0h
Yes
0 ÷ 1
Parameter“
tyPE
” of sub-menu “
bAr
” (display mode of the
bargraph),
0
- line;
1
- point
B1h
Yes
0 ÷ 6
Parameter “
FCol”
of sub-menu “
bAr
” (colours of the bargraph)
0
- bargraph off;
1
- option “
grEE
”;
2
- option “
rEd
”;
3
- option “
g r
”;
4
-option “
r g
”;
5
- option “
b g
”;
6
- option “
b r
”
B2h
Yes
-999 ÷ 9999 Parameter “
Lo C
” of sub-menu “
bAr
”, no decimal point
B3h
Yes
-999 ÷ 9999 Parameter “
Hi C
” of sub-menu “
bAr
”, no decimal point
B4h
Yes
0 ÷ 3
Parameter “
Pr1P
” of sub-menu “
bAr
” (relay's R1 thresholds'
markers):
0
– not showed;
1
- option ”
1St
”;
2
- option ”
2nd
”;
3
- option ”
botH
”;
B5h
Yes
0 ÷ 3
Parameter “
Pr2P
” of sub-menu “
bAr
” (relay's R2 thresholds'
markers):
0
– not showed;
1
- option ”
1St
”;
2
- option ”
2nd
”;
3
- option ”
botH
”;
B6h
Yes
0 ÷ 3
Parameter “
Pr3P
” of sub-menu “
bAr
” (relay's R3 thresholds'
markers):
0
– not showed;
1
- option ”
1St
”;
2
- option ”
2nd
”;
3
- option ”
botH
”;
B7h
Yes
0 ÷ 3
Parameter “
Pr4P
” of sub-menu “
bAr
” (relay's R4 thresholds'
markers):
0
– not showed;
1
- option ”
1St
”;
2
- option ”
2nd
”;
3
- option ”
botH
”;
B8h
Yes
0 ÷ 4
Parameter “
P1AL
” of sub-menu “
bAr
” (critical situation defined for
relay R1):
0
- no change;
1
- option “
morE
”;
2
- option “
LESS
”;
3
- option “
on
”;
4
- option “
oFF
”
B9h
Yes
0 ÷ 4
Parameter “
P2AL
” of sub-menu “
bAr
” (critical situation defined for
relay R2):
0
- no change;
1
- option “
morE
”;
2
- option “
LESS
”;
3
- option “
on
”;
4
- option “
oFF
”
BAh
Yes
0 ÷ 4
Parameter “
P3AL
” of sub-menu “
bAr
” (critical situation defined for
relay R3):
0
- no change;
1
- option “
morE
”;
2
- option “
LESS
”;
3
- option “
on
”;
4
- option “
oFF
”
BBh
Yes
0 ÷ 4
Parameter “
P4AL
” of sub-menu “
bAr
” (critical situation defined for
relay R4):
0
- no change;
1
- option “
morE
”;
2
- option “
LESS
”;
3
- option “
on
”;
4
- option “
oFF
”
1
- these registers are active only if device is equipped with active current output
2
- after writing to register no 20h the device responds with an “old” address in the message.
3
- after writing to register no 22h the device responds with the new baud rate.
4
- the value of the “
mbAc
” parameter is also connected to write to this register, so it is possible to block a writes, but
impossible to unblock writes via RS-485 interface, The unblocking of the writes is possible from menu level only.
5
-
the pairs of „
X
-
Y
” coordinates may be defined for any free point. The pair is “free” (it means that particular point is
not defined) if „
X
” coordinate of this point is equal 8000h. After writing both X and Y coordinate the point is defined
and used in calculation of result. The coordinates of any point can be changed at any time.
11.2.
TRANSMISSION ERRORS DESCRIPTION
If an error occurs while write or read of single register, then the device sends an error
code according to Modbus RTU specifications (example message no 1).
Error codes:
55