User manual - RATEMETER STI-N118
“mUL”
- multiplication coefficient. The value by which current measurement result is
multiply to recalculate it to desired unit (e.g. linear speed); can be set in range 0 to
999 999, value
0 is interpreted as 1 000 000.
“div”
- divide coefficient. The value by which current measurement result is divided to
recalculate it to desired unit (e.g. linear speed); can be set in range 0 to 999 999,
value
0 is interpreted as 1 000 000.
Parameters
“mul”
and
“div”
can be used to rescale of measurement value - setting
proper values recalculation of rotational speed to linear speed can be done automatically.
“ZEro t”
- maximum delay time of successive revolution. This parameter is defined with 0.1
second precision, in range 0 ÷ 39.9. If delay between successive revolutions are
longer than defined value, device displays „0” instead of current measurement.
If parameter
“ZEro t”
is set to value lower than parameter
“mEAS t”
, the
“mEAS t”
value is used as maximum delay to next revolution time
.
Expression „successive revolution” means the number of input pulses equals to
value of parameter
“PULSen”
.
“mEAS t”
- minimum
measurement window
width. The ”
measurement window
” means the
time period after which counted pulses are recalculated and displayed, and it is
similar to refresh rate. This parameter can be set with 0.1 sec. resolution in range
0 ÷ 19.9 sec.
If time delays between successive pulses are longer than
“mEAS t”
time, then
measurement window width is automatically fitted to input frequency.
7.3.3.
“ FiLtEr
” menu
This menu contains parameters referred to additional data (result) filtration.
“vALUE”
- this parameter sets filtration rate. It can be set to values from 0
(no filtration ) to 5 (strongest filtration – time window about 3.5 sec.).
“droP”
- this parameter defines minimum percentage change of measured value which turns
off (temporally) data filtration. Value of this parameter can be changed in range:
0 ÷ 199,9%. Main purpose of this parameter is acceleration of displaying data
changes, when sudden changes of instantaneous measurements value occur (while
parameter
“vALUE”
is set to high values).