User manual - RATEMETER STI-N118
Parameter
“
FrEq” (Hz)
Minimum duration
of low and high states
Permitted duty cycle
for max. frequency
”10”
5 ms
5%-95%
”15”
3,4 ms
5%-95%
”20”
2,5 ms
5%-95%
”30”
1,7 ms
5%-95%
”40”
1,3 ms
5%-95%
”50”
1 ms
5%-95%
”100”
500
m
s
5%-95%
”300”
167
m
s
5%-95%
”1 000”
50
m
s
5%-95%
”3 000”
33
m
s
10%-90%
”10 000”
33
m
s
33%-66%
”50 000”
not controlled
not controlled
Tab.7.1.Allowable settings of
“
FrEq”
parameter, and
related with them minimum durations of logical states
“PULSen”
- input divider, can be set in range 1 to 19 999. This parameter defines the
number of pulses per single revolution .
Value 0 is interpreted by firmware as
1
pulse/rev.
This parameter must be set correctly to let device works properly.
“PrEc”
-
precision of result displaying. There are available settings from
“0”
(precision of 1
unit) to
“0.00000”
(precision with 5 digits after decimal point)
“unit”
- The unit of rotational speed used for recalculation of current measurement can be
set to:
“SEC”
- result expressed in revolutions / second
“min”
- result expressed in revolutions / minute
“hour”
- result expressed in revolutions / hour
“F or P”
- kind of value shown on the display:
“Freq”
- frequency (rotational speed),
“PEriod”
- period (time per one revolution).
29