•   Connect the red lead to the anode and the black lead to the     

   cathode of the diode under testing.

Note:

•   The meter will show approximate forward voltage drop of the    

   diode.

•   If the lead connections is reversed, only '1' will be displayed.

Continuity Testing

•   Connect the black test lead to COM jack and the red to VΩmA  

   jack.

•   Set the rotary switch at the       range position.

•   Connect test leads across two points of the circuit under 

   testing.

•   If continuity exists (i.e. resistance less than about 50Ω), the    

   built-in buzzer will sound.

Note:

   If the input is an open circuit, the figure '1' will be displayed.

Measuring Capacitor

•   Connect the black test lead to Cx-jack and the red to Cx+ jack.

•   Set the rotary switch at the desired 20uF/200nF range position.

•   Before inserting capacitor under measurement into capacitance  

    testing socket be sure that the capacitor has  

 

 

    been discharged fully.

Measuring Battery

•   Load resistance 

1.5v range: 36Ω

   

 

9V range: 360Ω

   

 

12V range: 450Ω

•   The 1.5V battery is good when the display is more than 33mA.

•   The 9V battery is good when the display is more than 18mA.

•   The 12V battery is good when the display is more than 14mA.

MAINTENANCE

•   Before attempting to remove the battery door or open the case,  

    be sure that test leads have been disconnected from  

 

   measurement circuit to avoid electric shock hazard.

•   To avoid electrical shock, remove test leads from measurement  

    circuits before replacing the fuse. Replace fuses only with a    

   specified rating: F-200mA/250V fuse.

•   Replace the test leads if the lead is damaged or exposed in    

   any way. Only replace leads with the same type and 

   specification.

•   Use only moist fabric or a small amount of detergent, never 

    use chemical solutions for cleaning.

•   Do not use the meter before the back cover is properly closed  

   and screw secured.

•   Upon any signs of abnormality, stop operation immediately and  

   send the meter for maintenance by a qualified technician.