3.9   Programming of the set point

 

Push 

key P

, then release it. The programming mode is activated.

The 

upper display

 (PV) will visualise 

“SP 1”

 (Set Popint 1).

The 

lower display

 (SV) will indicate the 

actual programmed value

 of the heating block.

 

To modify the value, press 

5

 key to increase the value of the heating block

 

To modifythe value, press 

6

 key to decrease the value of the heating block

 

Once the desired value has been reached, by pushing 

key P

 it is possible to exit the programming mode.

NOTE:

A) The keys 

5

 and 

6

change the value 1°C at a time but if they are pressed for more than one second, the

value increases or decreases rapidly and, after two seconds in the same condition, the changing speed
increases in order to allow the desired value to be reached rapidly.
B) If no key is pressed for approx. 15 seconds, the display will return to normal functioning automatically.

4   Cooling down the heating block

Attention: Danger of burning (block and test piece).

Before transportation or touching the heating block the block has to be cooled down.

After switching on the instrument the installed fan starts to produce cool air, which leads to a sufficient thermic
decoupling of the enclosure.
For a quicker colling down of the heating block the fan can be switched to position 2 (S2) to produce a higher
amount of cooling air.

5   Recalibration

The test instrument, series TP18000E , have been adjusted and checked before delivery with measuring means,
which are based on nationally accepted standards.
Due to the DIN ISO 10 012 the instruments have to be checked depending on the area of application in
adequate periodical intervals. We recommend to send us the instruments in intervals of 12 months at the latest
or approx. 500 working hours to our plant for recalibration or readjustment.

6

   

Error signalling

Error

Reason

Action

- - - -

Probe interrupted

uuuu

The measured variable is under the probe’s
limits (under-range)

oooo

The measured variable is over  the probe’s limits
(over-range)

Verify the correct connection between probe
and instrument and then verify the correct
functioning of the probe

ErEP

Possible anomaly of the EEPROM memory

Push key “P”

In error conditions, the controller switches off the heating elements of the calibrator

7   Cleaning

We recommend  cleaning of the instrument with a slightly wet cloth using water and not abrasive cleaners or
solvents which may damage the instrument.

Summary of Contents for TP 18 600 E

Page 1: ...of the handle 4 3 2 Operating position 4 3 3 Adapter sleeves 4 3 4 Control of temperature sensors 4 3 5 Front panel description 4 3 6 Switching on of the calibrator 5 3 7 Indication of the block tempe...

Page 2: ...0 E correspond to the current state of technology This refers to measuring accuracy function as well as safe operation of the devices To grant safe operation competent and responsible behaviour of the...

Page 3: ...t for applications not expressly permitted by the above mentioned rule must adopt all the necessary protective measures The instrument CANNOT be used in dangerous environments flammable or explosive w...

Page 4: ...ry to use exactly adapted pockets The right bore is the outer dameter of the device under test 0 5 mm 3 4 Control of temperature sensors For the test of temperature sensors a separate temperature meas...

Page 5: ...ating cartridge heats the brass block up from ambient temperature to the adjusted set point value 3 7 Indication of the block temperature and set temperature The red 4 digit 7 segment LED PV shows con...

Page 6: ...ument the installed fan starts to produce cool air which leads to a sufficient thermic decoupling of the enclosure For a quicker colling down of the heating block the fan can be switched to position 2...

Page 7: ...s found without any fees or charge for SIKA except in the event of alternative agreements 10 Technical Data Regulation of the heating block with PID regulator Adjustment of the block temperature with...

Page 8: ...ensation Storage temperature 10 60 C Compliance ECC directive EMC 89 336 EN 61326 ECC directive LV 73 23 and 93 68 EN 61010 1 Dimensions TP18600E TP18600 28E TP18600 200E TP18600 28 200E heating block...

Reviews: