7 - Led SET

no function for temeprature calibrator

8 - Led AT/ST

no function for temeprature calibrator

9 – Display PV

It normally indicates the temperature of the heating block.
During programming it indicates SP1 (Set Point 1)

10 - Display SV

It normally indicates the active Set  value of the heating block

3.6   Switching on of the calibrator

 

After connection with the mains the fan starts immidiatly.

 

After switching on the mains power switch the controller starts with initialisation.
The upper display (PV) indicates tESt.
The lower display indicates the actual version of the controller e.a. rL 1.2

After 5 sec. the initialisation is complete and the standard mode is selected automatically.
The upper display (PV) indicates the actual heating block temperature
The lower display (SV) indicates the actual set value of the heating block
The installed electrical heating cartridge heats the brass block up from ambient temperature to the adjusted set
point value.

3.7   Indication of the block temperature and set temperature

The red, 4-digit, 7-segment LED (PV) shows continuously the temperature of the heating block. If the,  in the
green , 4-digit, 7-segment LED (SV), adjusted set temperature in the heating block is reached, the heating
energy radiated from the block is dispensed by short switching-on pulses of the heating cartridge, so that the
temperature inside of the block is kept constant.

3.8   Adjustment of the block temperature

The switch-on time of the heating is indicated by the red LED OUT1.
While during the heating-up period a permanent light shows the supply of heating energy. The flashing of the
LED shows that the temperature of the heating block is close to the adjusted set temperature, and therefore the
heating energy is now only supplied in short intervalls.
The control is carried out as a PID control.

In order to obtain good stability of the temperature, the cycle time “tcr1” has a low value with a very frequent
intervention of the control output.

Summary of Contents for TP 18 600 E

Page 1: ...of the handle 4 3 2 Operating position 4 3 3 Adapter sleeves 4 3 4 Control of temperature sensors 4 3 5 Front panel description 4 3 6 Switching on of the calibrator 5 3 7 Indication of the block tempe...

Page 2: ...0 E correspond to the current state of technology This refers to measuring accuracy function as well as safe operation of the devices To grant safe operation competent and responsible behaviour of the...

Page 3: ...t for applications not expressly permitted by the above mentioned rule must adopt all the necessary protective measures The instrument CANNOT be used in dangerous environments flammable or explosive w...

Page 4: ...ry to use exactly adapted pockets The right bore is the outer dameter of the device under test 0 5 mm 3 4 Control of temperature sensors For the test of temperature sensors a separate temperature meas...

Page 5: ...ating cartridge heats the brass block up from ambient temperature to the adjusted set point value 3 7 Indication of the block temperature and set temperature The red 4 digit 7 segment LED PV shows con...

Page 6: ...ument the installed fan starts to produce cool air which leads to a sufficient thermic decoupling of the enclosure For a quicker colling down of the heating block the fan can be switched to position 2...

Page 7: ...s found without any fees or charge for SIKA except in the event of alternative agreements 10 Technical Data Regulation of the heating block with PID regulator Adjustment of the block temperature with...

Page 8: ...ensation Storage temperature 10 60 C Compliance ECC directive EMC 89 336 EN 61326 ECC directive LV 73 23 and 93 68 EN 61010 1 Dimensions TP18600E TP18600 28E TP18600 200E TP18600 28 200E heating block...

Reviews: