Error And System Messages 

Display 

Meaning 

What to do? 

 

Low  battery power, device will only continue 

operation for a short period of time 

Replace battery 

Battery empty  

Replace battery 

 

Mains operation without battery: wrong voltage  Check  power supply, replace it when 

necessary 

Battery empty 

Replace battery 

Mains operation without battery: wrong voltage 

or polarity 

Check  power supply, replace it when 

necessary 

System error 

Disconnect battery and power supplies, 

wait shortly, then reconnect 

No display or 

confused 

characters, 

device does 

not react on 

keypress 

Device defective 

Return to manufacturer for repair 

Measured value above allowable range 

Check: pressure above 350mbar?  

-> measuring value to high 

Err.1 

Sensor defective 

Return to manufacturer for repair 

Measured value below allowable range 

Check: pressure below -10mbar?  

-> measuring value to high 

Err.2 

Sensor defective 

Return to manufacturer for repair 

Err.4 

Value is too low to be displayed, tara is set 

Check: display below -2000 (tara?)?  

Err.9 

Measured value far out of allowable range 

Check: pressure not within sensor range?  

Err.7 

System error 

Return to manufacturer for repair 

 

Summary of Contents for MH3160-07

Page 1: ...H Co KG Struthweg 7 9 34260 Kaufungen Germany 1 8 Tel 49 5605 803 0 Fax 49 5605 803 54 email info SIKA net web www SIKA net User s Manual Handheld Digital Precision Manometer MH3160 07 10 0 to 350 0 m...

Page 2: ...vice and sensors have to be handled with care don t throw hit etc Protect plugs and sockets from soiling 3 If the device is transported from a cold to a warm environment condensation may cause in a fa...

Page 3: ...ke sure that the operating voltage stated at the power supply is identical to the mains voltage Connections Connection for pressure tubes higher pressure lower pressure Interface Connection for el iso...

Page 4: ...e parameters can be changed with key 2 or key 5 Quit key 6 finishes the configuration and returns to standard measuring operation Unit Coice Of The Display Unit Choose the desired display unit the ref...

Page 5: ...e it when necessary System error Disconnect battery and power supplies wait shortly then reconnect No display or confused characters device does not react on keypress Device defective Return to manufa...

Page 6: ...nufacturer is capable to do efficient recalibration if necessary to get results of highest accuracy The Serial Interface By means of the serial interface and a suitable electrically isolated interface...

Page 7: ...6 Read min memory 200 Read min display range 7 Read max memory 201 Read max display range 12 Read ID number 202 Read display range unit 174 Clear min memory 204 Read display range decimal point 175 C...

Page 8: ...ce for the time of the power off delay The power off delay can be set to values between 1 and 120 min it can be completely deactivated Display 2 four digit LCDs 12 4mm high and 7 mm high for measuring...

Reviews: