6 Device Configuration
To change device settings, press
Set
(key 4) for 2 seconds. This will call the
configuration menu.
The parameters can be changed with
(key 2) or
(key 5).
Press
Set
again to go to the next setting.
Store
(key 6) finishes the configuration and returns to standard measuring operation.
Parameter
Values
Meaning
MENU
KEY
or
�
tYPE
ni.cr
J
S
T
N
B
E
Selection Thermocouple:
Type K: NiCr-Ni
Type J: Fe-CuNi
Type S: Pt10Rh-Pt
Type T: Cu-CuNi
Type N: NiCrSi-NiSi
Type B: Pt30Rh-Pt6Rh
Type E: NiCr-CuNi
rES
0.1°
Resolution
: 0,1°C or °F
1°
Resolution
: 1°C or °F
Unit
°C
Values in °C
°F
Values in °F
P.oFF
1-120
Auto Power Off
time in minutes
oFF
Auto Power Off deactivated
Out
oFF
Function of the output: No output function, lowest
power consumption
SEr
Output is serial interface
Adr.
01,11..91
Base
address
of interface (if Out = SEr)
OFS.
-10.0°C..10.0°C or
-18.0°F..10.0°F
The
offset of sensor
will be displaced by this value to
compensate for deviations in the Sensor or in the
measuring device.
oFF:
Zero point displacement is deactivated (=0.0°)
SCL.
-5.00..5.00 %
The measuring
scale of sensor
will be changed by
this factor [%] to compensate deviations of
temperature probe or measuring device.
oFF:
Correction factor is deactivated (=0.000)
SCF.
0.950..1.200
Correction value for surface measurement of sensor 1.
(related to
ambient temperature
of the measuring
device)
See chapter 9.3
Correction for surface measuring
oFF:
Surface correction factor is deactivated (=1.000)