Construction and function  

Technical changes reserved 

 

- 7 -

 

3

 

Construction and function 

Components: 

 

  Housing:  

The Housing consists of plastic and has the 

IP65 degree of protection.

 

 

 

Electrical connection:  

The electrical connection is made via 4

-pin 

plug M12x1.

 

 

 

Operation / flow indicator LED. 

 

 

Type plate with flow direction (marking) 

 

 

Process connection:  
The process connections are available in dif-
ferent sizes. 

 
 

Construction:  

The measuring tube with its earthing sleeves and electrodes passes 
through the housing and forms the external process connection of the 
VMZ.  
A magnetic field for the measurement process is generated inside the 
housing, which also contains the sensor and signal conditioning cir-
cuitry.  
The two stainless steel electrodes are located in the middle of the 
measuring tube between the earthing sleeves.  
The VMZ does not need any moving parts to make measurements. The inside of the measur-
ing tube is completely open, allowing the fluid to flow unhindered through the measuring 
tube. 

Function: 

The magnetic inductive flow sensor functions according to the induction principle. A DC volt-
age is generated by the movement of a conductor in a magnetic field: 
The measuring tube of the VMZ is located in a magnet-
ic field (B). 
An electrically conductive medium (V) flows through 
the measuring tube. The positive and negative charge 
carriers are oppositely deflected. 
A voltage (U) is generated at right angles to the mag-
netic field, which is picked up by the two electrodes.  
Thereby, the induced voltage is proportional to the av-
erage flow velocity of the liquid. 
The electronics of the VMZ converts the induced voltage into a flow-proportional frequency 
signal. 

 

 

 

Summary of Contents for induQ VMZ 030

Page 1: ...f the device is resold please provide the operating manual along with it SIKA Ba_VMZ_en 05 2016 Magnetic Inductive Flow Sensor Series VMZ Types VMZ 030 VMZ 081 VMZ 082 VMZ 153 VMZ 204 VMZ 205 VMZ 256...

Page 2: ...and off 11 6 3 Measuring operation 11 7 Maintenance and cleaning 12 7 1 Return shipment to the manufacturer 12 8 Disassembly and disposal 13 9 Technical data 14 9 1 Characteristics VMZ 14 9 2 Material...

Page 3: ...CAUTION Risk of injury This sign indicates dangers that cause personal injuries that can lead to health defects or cause considerable damage to property CAUTION Electric current This sign indicates da...

Page 4: ...n the front of the VMZ It contains the most important data and the connection dia gram for the electrical connection Example Fig 1 1 Delivery unpacking and accessories All units have been carefully ch...

Page 5: ...device supplied is only guaranteed by intended use The speci fied limits 9 Technical data may under no circumstances be exceeded Before installing the device check that the wetted materials of the de...

Page 6: ...must be aware of this operating manual and have access to it at all times The electrical connection should only be carried out by a fully qualified electrician General safety instructions In all work...

Page 7: ...el electrodes are located in the middle of the measuring tube between the earthing sleeves The VMZ does not need any moving parts to make measurements The inside of the measur ing tube is completely o...

Page 8: ...e equipment is depressurised and has cooled down SUITABLE TOOLS Use only suitable tools of the correct size 4 1 Installation instructions CAUTION Risk of malfunction due to external magnetic fields Ma...

Page 9: ...adjacent devices If adjacent devices are too close together operation of both devices may be impaired due to mutual interference 4 2 Assembly The VMZ is installed directly into the pipeline The compa...

Page 10: ...ize the electrical system before connecting the VMZ Connection and wiring Screw the coupling socket of the connection cable to the plug of the VMZ Tighten the knurled nut of the coupling socket with a...

Page 11: ...annot be switched on or off on its own Switching on and off is carried out by the applied supply voltage Switch on the supply voltage The red LED lights up permanently The VMZ is ready for use and goe...

Page 12: ...he operating company WARNING Risk of injury due to insufficient cleaning The operating company is responsible for all damages and harm of any kind in particu lar physical injuries e g caustic burns or...

Page 13: ...r to disassembly ensure that the equipment is switched off and is in a safe and de energised state the equipment is depressurised and has cooled down Disassembly Remove the electrical connectors Remov...

Page 14: ...10 0 12 5 Signal shape Square wave signal duty cycle 50 50 can be connected as PNP or NPN open collector Signal current 25 mA Electrical characteristics Supply voltage 24 VDC 15 or 12 VDC 15 Power co...

Page 15: ...nt Material Component wetted Housing ABS Measuring tube VMZ_ _ _ _ _DE_ _ _ _ _ _ VMZ_ _ _ _ _PE_ _ _ _ _ _ POM PVDF X Process connections VMZ_ _ _ _ _DE_ _ _ _ _ _ VMZ_ _ _ _ _PE_ _ _ _ _ _ POM PVDF...

Page 16: ...Technical data induQ Series VMZ 16 SIKA Ba_VMZ_en 05 2016 9 3 Pressure drop...

Page 17: ...l data Technical changes reserved 17 9 4 Dimensions VMZ DN 3 DN 8 DN 15 DN 20 Type L1 L2 D1 D2 VMZ 030 85 13 3 G B 3 VMZ 081 85 13 3 G B 8 VMZ 082 85 13 3 G B 8 VMZ 153 90 16 G B 14 VMZ 204 90 16 G1 B...

Page 18: ...Technical data induQ Series VMZ 18 SIKA Ba_VMZ_en 05 2016 VMZ DN 25...

Page 19: ...induQ Series VMZ Technical changes reserved 19 For your notes...

Page 20: ...05 2016 Sensors and Measuring Instruments Flow Measuring Instruments Test and Calibration Instruments SIKA Dr Siebert K hn GmbH Co KG Struthweg 7 9 D 34260 Kaufungen Germany 49 0 5605 803 0 49 0 5605...

Reviews: