Operation
i
ndu
Q
®
- Series VMM
- 40 -
© SIKA
•
Ba_VMM_us
•
09/2018
Direct simulation – current output *:
Using this function the requested value for
the current output can be defined.
For selecting the desired function within the
class please use the key
or
.
After selecting the function
Simulation di-
rect current output
, press the ENTER
key. The actual value for the current will be
displayed:
Example:
The current can be set within bounds of 0 to
23 mA. For changing the parameter please
refer to the description in Section "Input
window / modify a value".
* This parameter is of relevance only, if the pa-
rameter
preset of
is set to „direct“.
7.1.10
Self-test
The function class SELF-TEST includes all
functions relating to the self-test of the sen-
sor. The diagnostic functions of the display
electronics, which monitor the signals of the
sensor and proper functioning of the elec-
tronics and the software, are always active
and cannot be switched off.
Self-test on / off
:
The
Self-test on/off
function is intended
for enabling or disabling the monitoring
function of the field coil current. This func-
tion is useful for suppressing temperature
dependencies of the display electronics.
After selecting the function
Self-test
on/off
using the key
or
press the
ENTER key. The current setting will be dis-
played.
Set current
I = 12.50 mA