3 of 8

2760 DryLoc pH/ORP Preamplifier

‡ SIGNET

Use standard installation
hardware to connect the
submersible 2760 directly to
external equipment.

 1/2 in. NPT threads

The 8052-1 NPT Mount Junction box
connects to 

3

/

4

 in. pipe or conduit  and

provides convenient wiring termination.

The 8050-1 Universal Mount
junction box mounts flat onto a wall
or can be strapped to a post or
pipe.

1.

Insert any DryLoc electrode into base of the 2760 and turn
until keyed contacts are seated.

2.

Thread retaining cap over electrode and hand-tighten onto
2760.

A light coat of a non-petroleum based lubricant on the O-ring
will ease the installation.

Keep the DryLoc electrical interconnection between electrode
and preamplifier dry and clean at all times.

3.

Attach 

3

/

4

 in. watertight pipe to the top of the 2760.  Secure the

threaded connection to prevent any leakage.

For  additional defense against possible accumulation of
condensation at the back seal area of the 2760, fill the lower
3-4 inches (75-100 mm) of conduit or extension pipe with a
flexible sealant such as silicone.

2760 Submersible Sensor Assembly and Installation

94 mm/
3.7 in.

107 mm/
4.2 in.

Fill with

3 to 4 in.

of sealant

94 mm/
3.7 in.

106 mm/
4.2 in.

Fill with

3 to 4 in.

of sealant

Fill with

3 to 4 in.

of sealant

OR

OR

208 mm/
8.2 in.

55 mm/
2.16 in.

Fill with
75-100 mm
(3 to 4 in.)
of sealant

1.

2.

3.

Any combination of 2760 and electrode can be installed in a tank or vessel if conditions are within pressure and temperature
specifications.  The 2760 is able to withstand fluid temperatures up to 85°C, but the maximum tolerance of the electrodes varies.  Check
the specifications of each component before installation.

Main Contents

Previous View

pH/ORP Menu

Summary of Contents for DryLoc 2760

Page 1: ...terials CPVC Cable 4 6m 15 ft 3 cond shielded 22AWG Maximum cable 122 m 400 ft Shipping Weight 0 64 kg 1 41 lb Operating Temperature 0 C to 85 C 32 F to 185 F Storage Temperature 20 C to 85 C 4 F to 185 F Relative Humidity 0 to 95 non condensing without electrode connected Standards and Approvals CE Manufactured under ISO 9001 2000 for Quality and ISO 14001 1996 for Environmental Management Descri...

Page 2: ...the threads to secure the assembly 5 Use 3 4 in rigid or flexible conduit to protect the cable from damage 6 If the cable will be extended beyond the standard 15 ft supplied with the 2760 use a junction box to terminate the wires Note Do not extend the 2760 cable more than 120 m 400 ft In line Assembly and Installation 178 mm 7 in 107 mm 4 2 in 3 4 in Conduit customer supplied Fitting 2764 Differe...

Page 3: ... 4 in watertight pipe to the top of the 2760 Secure the threaded connection to prevent any leakage For additional defense against possible accumulation of condensation at the back seal area of the 2760 fill the lower 3 4 inches 75 100 mm of conduit or extension pipe with a flexible sealant such as silicone 2760 Submersible Sensor Assembly and Installation 94 mm 3 7 in 107 mm 4 2 in Fill with 3 to ...

Page 4: ... 2756 DI 3 2757 Standard Flat DryLoc 3 2754 3 2754 HF 3 2755 Connect this terminal to Earth GND to reduce electrical interference White Temp Green Temp Blue Iso Gnd Brown mV input Shield Black V Red V 12 11 10 9 8 7 6 5 18 17 16 15 14 13 12 11 16 15 14 13 12 11 10 9 3 8750 1 3 8750 2 3 8750 3 ProcessPro pH ORP Transmitter Any of these electrodes can be connected to the GF SIGNET ProPoint and Proce...

Page 5: ...RTD 3 2766 1 3KΩ RTD 3 2764 2 PT1000 RTD 3 2766 2 PT1000 RTD 3 2764 3 300Ω RTD 3 2766 3 300Ω RTD DryLoc Preamplifier 3 2760 1 3 2760 2 White Temperature Green Temperature Blue mV Reference Shield Earth Ground Black V Red V Brown Active mV pH ORP Combination ORP Redox Electrodes 3 2775 10KΩ ID 3 2777 10KΩ ID 3 2775 1 no Temp Comp 3 2777 1 no Temp Comp 3 2755 10KΩ ID 3 2757 10KΩ ID 3 2757 WT 10KΩ ID...

Page 6: ...6 of 8 2760 DryLoc pH ORP Preamplifier SIGNET Notes Main Contents Previous View pH ORP Menu ...

Page 7: ...7 of 8 2760 DryLoc pH ORP Preamplifier SIGNET Notes Main Contents Previous View pH ORP Menu ...

Page 8: ...process connection 3 2764 2 159 000 944 Differential pH Electrode PT1000 RTD 1 in NPT process connection 3 2764 3 159 000 945 Differential pH Electrode 300Ω RTD 1 in NPT process connection 3 2766 1 159 000 949 Differential pH Electrode 3KΩ RTD 1 in NPT process connection 3 2766 2 159 000 950 Differential pH Electrode PT1000 RTD 1 in NPT process connection 3 2766 3 159 000 951 Differential pH Elect...

Reviews: