
37
Signet 8900 Multi-Parameter
SENSOR:
Select to perform calibration at the sensor via 2750 EasyCal.
INSTRUMENT: Select to perform calibration at the 8900 via the steps below.
If Cal at Instrument is selected, the following menu items appear:
Place sensor in any pH buffer. Allow suffi cient time for sensor output to stabilize, then
enter value. Maximum offset allowed: ±1.3 pH.
Place sensor in any buffer that is different by at least 2 pH units from Std value, then enter
value. Maximum offset allowed: ±1.3 pH.
pH1 Cal at:
Sensor >
Adjust the temperature measurement from the pH sensor to match a reliable reference.
Maximum offset allowed is ± 20 °C
If pH~ Cal at Instrument is selected, reset Standard, Slope and Temperature to factory
calibration.
13.2 Calibration: If Channel type = pH
ORP1 Cal at:
Sensor >
13.3 Calibration: If Channel type = ORP
Set pH1 Std:
>
Set pH1 Slope:
>
Set Temperature
T1: >
Reset pH1 to
Factory Cal: >
SENSOR:
Select to perform calibration at the sensor via 2750 EasyCal.
INSTRUMENT: Select to perform calibration at the 8900 via the steps below.
If Cal at Instrument is selected, the following menu items appear:
Place sensor in any ORP test solution. Allow suffi cient time for sensor output to
stabilize, then enter value. Maximum offset allowed: ±2900 mV.
Place sensor in any buffer that is different by at least 30 mV from Std value, then enter
value. Maximum offset allowed: ±2900 mV.
If ORP~ Cal at Instrument is selected, reset Standard and Slope to factory calibration.
Set ORP1 Std:
>
Set ORP1 Slope:
>
Reset ORP1 to
Factory Cal: >