Signametrics
46
directory.
e
verall analog performance for that DMM. You must not alter this file unless you are performing an external
a
.
s. 1
st
entry is Offset the 2
nd
is gain parameters
86
-37.
3.
.8
Subsequent lines: 1
st
entry is Offset the 2
nd
is gain, 3
rd
freq. comp
.303
; starting with the 330mV range, and last line is for the 250V range.
c
; IAC 3.3mA to 2.5A ranges, offset and gain
1.6
.0
.001848
.0
…
.
.
The
During
initi
driver reads information from the DMM hardware such as serial number, DMM type, and allocated
O space. Next it reads the corresponding calibration information from the
SM40CAL.DAT
file and makes sure
od and the software. If all is checked OK, it initialize
SM2044.exe
Visual Basic DMM control panel executable
Msvcrt.dll
System file. Installs in your
C:\WINDOWS\SYSTEM
Windrvr.vxd
Win98/95 Virtual Device Driver. Installs by ‘setup’ in your
C
:\WINDOWS\SYSTEM\VMM32
directory.
Windrvr.sys
Win NT Virtual Device Driver. Installs by ‘setup’ in your
C
:\WINNT\SYSTEM32\DRIVERS
directory.
Install.doc
Installation instructions in MS Word
5.1.1 The SM40CAL.DAT file
The
SM40CAL.DAT
file contains calibration information for the specific DMM is it included. It determines th
o
calibration of the DMM. This file may contain multiple records for more than one DMM. Each record starts with
header line, followed by calibration data.
card_id 10123
type 2044 calibration_date 06/15/1999 ; The identification should match the last digits of the SN
ad
; A/D compensation
72.0 20.0
dc
; VDC 330mV, 3.3V, 33V, 330V range
v
-3
.0 0.99961
0 .999991
0 0.999795
-8
-8 1.00015
; VAC 1st line - DC offset.
vac
5
0.84
1.015461
23
0.0043
1.0256
23
0.0
1.02205
0
0.0
1.031386
0
idc
; IDC 3.3mA to 2.5A ranges. 1
st
entry is offset, 2
nd
is gain parameter
-1450.0 1.00103
-176.0 1.00602
-1450.0 1.00482
-176.0 1.0
ia
1.02402
1.03357
0
1.69 1.00513
0.0 1.0142
2w-ohm ; Ohms 33, 330, 3.3k,...,330Meg ranges, offset and gain
12700.0
1.002259
;in the SMX2040, the 1
st
and last lines are placeholders
1256.0
1.002307
110.0
1.002665
0.0
1.006304
0.0
1.003066
.0
1
0
0
0.995664
0.0 1.00030
first line identifies the DMM and the calibration date. The "card-id" is stored in ROM on each DMM.
alization the
I/
there is compatibility between the hardware, the calibration rec
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com