background image

12

QUIRC User Guide

>

ll

h

file

i

Read DSP code .lod file

ll reads a DSP code binary file into the Sparcstation memory.

>

dl

hi

Download DSP code

dl sends the DSP code to the DSP electronics, and then causes the DSP to start

executing this code. dl also asks the DSP to recalibrate its A/D converter, which takes
about 2 seconds.

>

du

h

file

i

Download Utility board code

du loads the utility board code. This command should be run after the DSP has been

initilalized with the

dl

or

df

commands above. The utility board code is necessary to

perform filter wheel and pupil slide motions.

Example:

du util

8.2

Commands for setting up and taking exposures

Qcdcom takes exposures and writes data files according to various parameters which you can set.
The basic sequence that qcdcom follows is to

1. Reset the chip

2. Do dummy read to stabilize array

3. Perform first data read of double-correlation pair, store image in memory

4. Turn off output amplifiers

5. Open the shutter and wait for a predetermined exposure time

6. Close the shutter

7. Turn on output amplifiers

8. Do dummy read to stabilize array

9. Perform second data read, subtract the first from it, store in memory

10. Write the contents of memory into a FITS format data file

11. Put the controller into reset idle mode

These actions are performed by the fundamental command

go

, which is described below. First, the

commands which tailor the sequence followed by

go

are given.

Not all of these steps may be taken, and the actual sequence followed depends on how the automatic
actions are set, as described below. There are a number of ways to set the parameters which govern
what qcdcom does. Several commands are available which can be used to set integration time, object
name, object type (a FITS header item), and default automatic actions. None of these does anything
more than setting parameters to be used later—only

go

takes an exposure.

Summary of Contents for QUick Infrared Camera

Page 1: ...Observer s Guide for QUIRC K W Hodapp J L Hora M R Metzger University of Hawaii October 30 1997...

Page 2: ......

Page 3: ...ith QUIRC The qcdcom program controls taking exposures and writing data in FITS format to disk operates the moving parts of the instrument such as the shutter filter wheel and pupil mask communicates...

Page 4: ...ation differences for integration times less than 1 second Therefore short exposures should be avoided particularly when exposing dome flats it is far better to dim the lights with the domelight dimme...

Page 5: ...elow assume a PSF of 0 5 arcsec FWHM Table 2 QUIRC Sensitivity UH 2 2m Telescope Filter Point SRC Mag Mag arcsec2 Jy ADU 1 sec J 18 6 17 9 1 052e 6 H 17 8 17 1 8 295e 7 K 16 2 15 5 9 622e 7 H K 18 6 1...

Page 6: ...ted image display capability A separate program should be used to display the data One option is to use the viewfits program vf This is a display program developed by Tony Denault at the IRTF that dis...

Page 7: ...ode is running The command to download the current standard DSP code is the following df quirc Occasionally the download will fail if the camera is in some undetermined state and a reset will need to...

Page 8: ...rash for any reason the parameters it reads in may be old and you may need to set them again Typically only resetting the file number is necessary 3 3 Taking exposures For each object you are observin...

Page 9: ...telescope and guide probe and turns the guider back on On the 88 this communicates with Atlas and MiniOtto over the ethernet to perform the guider movements and on CFHT commands are sent over a seria...

Page 10: ...6 H K dual 1 8 m 2 794 20 6 6 1 7 H2BP23 2 12 m 0 914 2 1 7 1 8 2260BP60 2 260 m 2 794 20 6 8 2 1 1560BP120 1 56 m 9 2 2 1710BP120 1 71 m 10 2 3 1 989BP50 1 989 m 11 2 4 1 506 60 1 506 m 12 2 6 1 107...

Page 11: ...th auto or the shutter will not open Also note that you should always do only one filter DARK followed by exactly one filter BACK always in pairs otherwise the filter wheel will get lost and require a...

Page 12: ...chop image and displayed in vf Therefore the observer can take focus frames acquire and center the object etc without saving all the images as separate data files This avoids filling the disk with use...

Page 13: ...P initialization commands In general you will need to initialize the DSP processor in the Leach electronics after power up or after a DSP reset because its cold reset state is not appropriate for runn...

Page 14: ...rform first data read of double correlation pair store image in memory 4 Turn off output amplifiers 5 Open the shutter and wait for a predetermined exposure time 6 Close the shutter 7 Turn on output a...

Page 15: ...e double correlated read on off Controls whether the device is read out before and after the shutter opening or only after If reading before and after the difference between the two are taken and this...

Page 16: ...o stay open slightly longer The exposure time recorded in the header however will reflect the true amount of time the shutter was open filter hnum commandi Set filter This command selects the filter s...

Page 17: ...fp q940811 go h num i Start exposure Go is the command to take an exposure As described above according to the param eters which are set go may reset the chip open the shutter and close it a predeterm...

Page 18: ...resulting image to a disk file set hparami n Set parameter Certain program parameters can be set or unset with this command To set a param eter for example save one issues the command set save To uns...

Page 19: ...direct access to the DSP computer hall hnloopi Read Hall sensors read the Hall effect sensors nloop times The filter wheel 1 sensor is displayed and depending on the position of the switch either the...

Page 20: ...en data is used as a seed for a random number generator and 65536 words of random data are sent per loop diag l3 hcnt fr cg data i Timing board loopback test Loopback test with timing board in place M...