QUIRC User Guide

>

name

h

objname

i

Set object name

name is used to set the object name which will be written to the FITS header.

>

pupil

h

num command

i

Set pupil position

This command sets the pupil position. If no pupil number is given, a list of available

pupils are given and the current pupil position is shown. If the pupil number is different
from the current, the pupil slide is moved to the new position. Note that the positions are
separated by 7500 steps of the motor, so it takes many minutes to move to a new position.
The pupil can be moved to the reference position by typing

pu home

.

The command

pu switch

will report on the position of the pupil/filter wheel 2

switch. The same motor controller is used for these motors, and only one can be in
operation at any time, selected by the switch on the utility (black) box on the QUIRC
dewar. This command tells whether it is in the PUPIL or FILTER position.

For fine adjustment of the pupil position, the commands

pu FO

and

pu BA

will

move the pupil 10 or -10 steps, respectively. This can be used to move the pupil by small
amounts near its nominal position.

>

rc

h[

Xnx Xny

]i

read array

rc reads out the array, and if the auto-write flag is set, rc will write the resulting image

to a disk file.

>

set

h

param

i

[

n

]

Set parameter

Certain program parameters can be set or unset with this command. To set a param-

eter, for example “save", one issues the command

set save

. To unset, the command

set nosave

is used. Alternatively, one can type

set save 1

and

set save 0

to turn it on or off. Below are the available parameters:

clobber – If on, existing files will be automatically overwritten. If off, the program
will ask for confirmation from the user before overwriting files.

chop – Turns chopping on or off. Also can be set via the

auto

command.

beep – Causes the program to beep after each detector readout, and other operations.
Can be annoying.

vfout – Directs qcdcom to send commands to vf to display the most recent image.

save – Turns auto save on or off. Also can be set via the

auto

command.

longshut – Sets the long shutter mode. In this mode, the shutter is opened before
the initial reset of the observation is done, and keeps it open until after the final
read is completed. Because of the array read time, the minimum integration time
is about 11 seconds. This mode can be used if one wants to avoid the background
from the warm shutter when taking exposures of faint fields. Note that if any bright
objects are in the field the detector will likely saturate on those sources.

>

shutter

h

(open|close)

i

Commands the specified shutter action.

>

status

hi

Show status

status shows the current parameters.

Summary of Contents for QUick Infrared Camera

Page 1: ...Observer s Guide for QUIRC K W Hodapp J L Hora M R Metzger University of Hawaii October 30 1997...

Page 2: ......

Page 3: ...ith QUIRC The qcdcom program controls taking exposures and writing data in FITS format to disk operates the moving parts of the instrument such as the shutter filter wheel and pupil mask communicates...

Page 4: ...ation differences for integration times less than 1 second Therefore short exposures should be avoided particularly when exposing dome flats it is far better to dim the lights with the domelight dimme...

Page 5: ...elow assume a PSF of 0 5 arcsec FWHM Table 2 QUIRC Sensitivity UH 2 2m Telescope Filter Point SRC Mag Mag arcsec2 Jy ADU 1 sec J 18 6 17 9 1 052e 6 H 17 8 17 1 8 295e 7 K 16 2 15 5 9 622e 7 H K 18 6 1...

Page 6: ...ted image display capability A separate program should be used to display the data One option is to use the viewfits program vf This is a display program developed by Tony Denault at the IRTF that dis...

Page 7: ...ode is running The command to download the current standard DSP code is the following df quirc Occasionally the download will fail if the camera is in some undetermined state and a reset will need to...

Page 8: ...rash for any reason the parameters it reads in may be old and you may need to set them again Typically only resetting the file number is necessary 3 3 Taking exposures For each object you are observin...

Page 9: ...telescope and guide probe and turns the guider back on On the 88 this communicates with Atlas and MiniOtto over the ethernet to perform the guider movements and on CFHT commands are sent over a seria...

Page 10: ...6 H K dual 1 8 m 2 794 20 6 6 1 7 H2BP23 2 12 m 0 914 2 1 7 1 8 2260BP60 2 260 m 2 794 20 6 8 2 1 1560BP120 1 56 m 9 2 2 1710BP120 1 71 m 10 2 3 1 989BP50 1 989 m 11 2 4 1 506 60 1 506 m 12 2 6 1 107...

Page 11: ...th auto or the shutter will not open Also note that you should always do only one filter DARK followed by exactly one filter BACK always in pairs otherwise the filter wheel will get lost and require a...

Page 12: ...chop image and displayed in vf Therefore the observer can take focus frames acquire and center the object etc without saving all the images as separate data files This avoids filling the disk with use...

Page 13: ...P initialization commands In general you will need to initialize the DSP processor in the Leach electronics after power up or after a DSP reset because its cold reset state is not appropriate for runn...

Page 14: ...rform first data read of double correlation pair store image in memory 4 Turn off output amplifiers 5 Open the shutter and wait for a predetermined exposure time 6 Close the shutter 7 Turn on output a...

Page 15: ...e double correlated read on off Controls whether the device is read out before and after the shutter opening or only after If reading before and after the difference between the two are taken and this...

Page 16: ...o stay open slightly longer The exposure time recorded in the header however will reflect the true amount of time the shutter was open filter hnum commandi Set filter This command selects the filter s...

Page 17: ...fp q940811 go h num i Start exposure Go is the command to take an exposure As described above according to the param eters which are set go may reset the chip open the shutter and close it a predeterm...

Page 18: ...resulting image to a disk file set hparami n Set parameter Certain program parameters can be set or unset with this command To set a param eter for example save one issues the command set save To uns...

Page 19: ...direct access to the DSP computer hall hnloopi Read Hall sensors read the Hall effect sensors nloop times The filter wheel 1 sensor is displayed and depending on the position of the switch either the...

Page 20: ...en data is used as a seed for a random number generator and 65536 words of random data are sent per loop diag l3 hcnt fr cg data i Timing board loopback test Loopback test with timing board in place M...

Reviews: