5.1.6 Safety
Devices
Apart from the mentioned passive safety devices due to the solid design, the
following active precautions are in place for your safety:
5.1.6.1 Lid Lock and Lid Closing Device
The centrifuge can only be started when the lid is properly closed. The electrical
locks must be locked. The lid can only be opened when the rotor has stopped. If
the lid is opened by way of the emergency release system during operation, the
centrifuge will immediately switch off and decelerate brakeless. If the lid is open,
the drive is completely separated from the mains power supply, i.e. the centrifuge
cannot be started (see 6.2.5 "Emergency Lid Release", page 42).
5.1.6.2 Standstill
Monitoring
Opening of the centrifuge lid is only possible if the rotor is at a standstill. This
standstill is checked by the microprocessor.
5.1.6.3 System
Check
An internal system check monitors the data transfer and sensor signals with
regard to plausibility. Errors are detected with extreme sensitivity and displayed
as error messages with a number on the speed display (see 6.2.6 "Error Codes",
page 43).
5.1.6.4 Ground
Wire
Check
For the ground wire check, there is an equipotential bonding screw on the rear
panel of the centrifuge. A ground wire check can be carried out by authorized and
specialized personnel using a suitable measuring instrument (see 4.2.2
"Connection“, page 24). Please contact the head of our service team (see 6.2.7
"Service Contact", page 43).
Operating Manual SIGMA 1-6P
page 28 of 67
05/2007
Summary of Contents for 1-6P
Page 1: ...Laboratory Centrifuge 1 6P Operating Manual From Serial No 119956 Please retain for later use...
Page 2: ......
Page 4: ......
Page 6: ......
Page 12: ......
Page 14: ......
Page 54: ...Operating Manual SIGMA 1 6P page 54 of 67 05 2007 Abb 11 1 Drehzahl Schwerefeld Diagramm...
Page 56: ......
Page 58: ......
Page 60: ......
Page 64: ......