4.2 Installation
4.2.1 Installation
Site
All the energy supplied to the centrifuge is converted into heat and emitted to
ambient air.
•
Ensure sufficient ventilation.
•
Keep a safety distance of at least 30 cm from the wall so that the vents in the
centrifuge remain fully effective.
•
Do not position the centrifuge near heat generators.
•
Avoid direct sunlight (UV radiation).
•
The table must be stable and have a solid, even surface.
)
Attention!
During transport from cold to warmer places, condensation will collect inside the
centrifuge. It is important to allow sufficient time for drying before the centrifuge
can be used again.
4.2.2 Connection
The operating voltage on the name plate must correspond to the local
supply voltage!
SIGMA laboratory centrifuges are units of safety class I in accordance with DIN
VDE 0700 and have a three-wire power cord (2.5 m) and a shockproof right-
angle plug.
On the back, next to the mains power input, there is an additional ground wire
connector (see Fig. 5.2, page 25) where a separate ground wire can be
connected. This ensures that the admissible leakage current cannot shock
anyone if a problem with the protective ground wire system were to occur. Only
authorized and specialized personnel are permitted to connect the unit. Please
contact the head of our service team (see 6.2.7 "Service Contact", page 43).
4.2.3
Fuses/Emergency Circuit Breaker on Site
Typically, the centrifuges must be protected with 16 A fuses of class "G".
An on-site emergency circuit breaker to disconnect the centrifuge from the power
supply in the event of a malfunction is recommended. This switch should be
located away from the centrifuge, preferably outside the room where the
centrifuge is located or at the exit of this room.
Operating Manual SIGMA 1-6P
page 24 of 67
05/2007
Summary of Contents for 1-6P
Page 1: ...Laboratory Centrifuge 1 6P Operating Manual From Serial No 119956 Please retain for later use...
Page 2: ......
Page 4: ......
Page 6: ......
Page 12: ......
Page 14: ......
Page 54: ...Operating Manual SIGMA 1 6P page 54 of 67 05 2007 Abb 11 1 Drehzahl Schwerefeld Diagramm...
Page 56: ......
Page 58: ......
Page 60: ......
Page 64: ......