background image

 

SIGLENT 

SPD3303S/3303D  Quick  Start    27

 

   

Command format 

INSTrument?

 

Description 

Query the current operating channel 

 

Example 

INSTrument?

 

Return Info 

CH1 

 

5.    MEASure

 

Command format 

MEASure: CURRent? < CH1|CH2>

 

Description 

Query current value for specified channel, if there is no 

specified channel, query the current channel.

 

Example 

MEASure: CURRent? CH1

 

Return Info 

3.000 

 

Command format 

MEASure: VOLTage? < CH1|CH2>

 

Description 

Query voltage value for specified channel, if there is no 

specified channel, query the current channel.

 

Example 

MEASure: VOLTage? CH1

 

Return Info 

30.000 

 

Command format 

MEASure: POWEr? < CH1|CH2>

 

Description 

Query power value for specified channel, if there is no 

specified channel, query the current channel.

 

Example 

MEASure: POWEr? CH1

 

Return Info 

90.000 

 

 

 

Summary of Contents for SPD3303D

Page 1: ...Quick Start SPD3303S 3303D Programmable DC Power Supply V1 0 2013 SIGLENT TECHNOLOGIES CO LTD...

Page 2: ...publication replaces all previous corresponding material 3 SIGLENT reserves the right to change the specifications and price of the product 4 Any way of copying extracting and translating the content...

Page 3: ...ply AC Input Voltages 100V 110V 220V 230V 10 50 60Hz Use proper fuse The fuse types 100V 110V T6 3A 250V 220V 230V T3 15A 250V Make sure to use the correct type of fuse before turning on the instrumen...

Page 4: ...ge Please keep proper ventilation and check the fan and air vents regularly when using the instrument Operate condition Location indoor no strong light almost no Interfering pollution Comparative humi...

Page 5: ...icates direct injury or hazard that may happen WARNING Indicates potential injury or hazard that may happen CAUTION Indicates potential damage to the instrument or other property that may happen Symbo...

Page 6: ...D3303S 3303D 4 3 TFT color LCD display with content of 16M Three independent outputs two of which are adjustable and the total power up to 195W Four kinds of input voltage values includes 100V 110V 22...

Page 7: ...7 Output check 8 Chapter 2 Control panel operation 9 2 1 Output summary 10 2 2 CH1 CH2 Independent Output 11 2 3 CH3 Independent mode 12 2 4 CH1 CH2 Series mode 13 2 5 CH1 CH2 Parallel mode 15 2 7 Ti...

Page 8: ...contents of the shipment have been completely checked and the instrument has passed both electrical and mechanical tests The consigner or carrier will be responsible for damages to the instrument res...

Page 9: ...e DIP Switch at the rear panel according to the actual demand Warning To switch to the wanted power voltage please disconnect the power cord first Electrical Check Please use the power cord provided a...

Page 10: ...Area 10 Power Switch 3 Model 11 CH1 Output Terminal 4 System parameter configuration button 12 Ground Terminal 5 Multi function knob 13 CH2 Output Terminal 6 Fine Adjust button 14 CV CC indicator lig...

Page 11: ...setting interface Press the button longer to lock it and shorter to turn to system information interface Buttons for controlling the channel Press the button to turn on off all the channels Press the...

Page 12: ...on front panel The output terminals of CH1 CH2 and CH3 include positive terminal and negative terminal and a common ground for CH1 and CH2 additional Each channel has its own logo The operation detail...

Page 13: ...annel logo 3 Operating mode logo The corresponding logo will display when working in CV or CC mode 4 LOCK logo It will display when the LOCK is turned on 5 USB logo It will display when there is a USB...

Page 14: ...panel Description 1 warning message 2 The DIP switch of the AC power and its identification 3 The description of the AC input voltage 4 AC power socket 5 The fan air vents 6 CE certification mark 7 U...

Page 15: ...values of all channels are not zero 2 Check the voltage output of CH1 CH2 Turn on CH1 CH2 and the instrument work in CV mode Check whether the voltage value could be changed from 0V to 32V 2 Current o...

Page 16: ...tion of SPD3303S 3303D control panel will be introduced in detail to giving you an all around understanding of it which will eventually lead to easier work Brief introduction Output summary CH1 CH2 in...

Page 17: ...e that of the single channel In the series mode the voltage value is twice that of the single channel Constant voltage current In the constant current mode independent or tracking mode the current val...

Page 18: ...egative terminals of CH1 CH2 3 Set voltage and current value of CH1 CH2 Firstly move the cursor to select the wanted parameter voltage current and then revolve the multi function knob to change the co...

Page 19: ...3V 5V 3A Operation steps 1 Connect the load to the positive and negative terminals of CH3 on the front panel 2 Select the wanted voltage value by moving CH3 DIP switch 3 Open output Press ON OFF butt...

Page 20: ...eries mode and the indicator light turns bright immediately 2 Connect the load to the positive terminal of CH2 and the negative terminal of CH1 3 Turn on CH2 and then adjust its current value by revol...

Page 21: ...SIGLENT 14 SPD3303S 3303D Quick Start Note You can identify the current working state CC or CV by referring to indicator light of CH1 CH2 red means CV green means CC...

Page 22: ...steps 1 Press the PAR button to start Parallel mode and the indicator light turns bright immediately 2 Connect the load to the positive and negative terminal of CH1 3 Turn on CH1 and then set the outp...

Page 23: ...SIGLENT 16 SPD3303S 3303D Quick Start...

Page 24: ...FILE CHOICE by revolving the multi function knob 4 Select OPEN CHOICE by moving the cursor 5 Move the cursor to STORE and then press it to save the current setup Recall setup Operation steps 1 Press...

Page 25: ...consecutive output and the longest time of each group is 10000s Setup steps 1 Press CH1 CH2 to select the wanted channel 2 Press TIMER to enter the Timer Setup interface and the indicator light turns...

Page 26: ...n button 2 Revolve the multi function knob to make theTimer state to ON 3 Press the knob to start theTimer Note If press the TIMER button when the timer is running then you can observe the decreasing...

Page 27: ...button to enter Waveform Display interface and the indicator light turns bright immediately 3 Press CH1 CH2 ON OFF button to turn on CH1 CH2 channel the corresponding indicator light turns bright imm...

Page 28: ...SIGLENT SPD3303S 3303D Quick Start 21 2 9 Version information Press LOCK button quickly to enter Version Information interface which is shown below...

Page 29: ...ce 1 Open the EsayPower software after USB line having been connected perfectly and make sure the instrument is connected with the software correctly 2 Click Version and then choose Upgrade in the dro...

Page 30: ...3D Quick Start 23 Figure 2 Figure 3 4 As it shows in figure 4 click Upgrade button to start upgrading The upgrade is finished when the progress bar complete and the instrument will run the version aft...

Page 31: ...dure also can be used if the method above is failed Specific steps are below 1 Press the knob before powering on the instrument and the turn on the instrument and it will enter the guide procedure mod...

Page 32: ...n installed with EasyPower software or NI Measurement Automation 3 1 Command list 1 IDN 2 SAV 3 RCL 4 INSTrument CH1 CH2 5 INSTrument 6 MEASure CURRent 7 MEAsure VOLTage 8 MEASure POWEr 9 SOURce CURRe...

Page 33: ...are version Example Siglent SPD3303S SPD00001130025 1 01 01 02 2 SAV Command format SAV name Description Save current state in nonvolatile memory with the specified name Example SAV 1 3 RCL Command fo...

Page 34: ...d channel query the current channel Example MEASure CURRent CH1 Return Info 3 000 Command format MEASure VOLTage CH1 CH2 Description Query voltage value for specified channel if there is no specified...

Page 35: ...ce CURRent SOURce CH1 CH2 Description Query the current value of the selected channel Example CH1 CURRent Return Info 0 500 Command format SOURce VOLTage value SOURce CH1 CH2 Description Set voltage v...

Page 36: ...t 1 Series 2 Parallel Description Select operation mode Example OUTPut TRACK 0 Command format OUTPut WAVE SOURce state SOURce CH1 CH2 state ON OFF Description Turn on off the Waveform Display function...

Page 37: ...nd format TIMEr SOURce state SOURce CH1 CH2 state ON OFF Description Turn on off Timer function of specified channel Instruction The command works effectivly only when secnum starts from 1 Example TIM...

Page 38: ...nfo into a binary The state correspondence relationship is as follow Example SYSTem STATus Return info 0x0224 Bit NO Corresponding state 0 0 CH1 CV mode 1 CH1 CC mode 1 0 CH2 CV mode 1 CH2 CC mode 2 3...

Page 39: ...H3 Power coefficient 5mV Load coefficient 15mV Ripple and noise 2mVrms 5Hz 1MHz Tracking Track error 0 5 10mV of Master No Load operation Parallel coefficient Line 0 01 3mV Load 0 01 3mV rating curren...

Page 40: ...cord 30M or above DC 500V Operating Indoor environment Altitude 2000 m Environment temperature 0 40 Comparative humidity 80 Installation level Pollution level 2 Storage Environment temperature 10 70 e...

Page 41: ...not 0 Answer 3 Yes because when current output load is higher than the limited current value the working mode turns to CC from CV Question 4 How to deal with the upgrade failed Answer 4 please do it a...

Page 42: ...S 3303D Quick Start 35 Contact SIGLENT MTR Add 3 F Building 4 Antongda Industrial Zone 3rd Liuxian Road 68 District Baoan District Shenzhen P R CHINA Tel 86 755 36615186 E mail sales siglent com http...