CP6000/CP4000
16 Current Probe Instructions
When the input current continuously surpasses the maximum input range, the
self-protection will be activated by the heating of the sensor and cause wrong
output. Please stop the current input and wait for full cool down before next
operation.
The protection circuit will be mistakenly activated by the high temperature even
when the continuous current under test is below the max input.
When the connect input surpass the max input range current and activate the
protection function too often, the device may be damaged.
You must open the entire part through switch controller.
At the lock state, please do not press the entire part as shown below.
Preparation before testing
Prepare the high frequency current probe CP6000 series, adapter and
oscilloscope
Power up the CP6000 probe and the green LED power indicator will be lighted.
Set the oscilloscope: Ground the measuring mode, zero set the oscilloscope
and turn the oscilloscope mode to DC mode.
Choose the proper range according to the current under test. The default
setting of the probe is large current range
Degaussing and Zero setting
Connect the CP6000 with oscilloscope ( Make sure the input impedance of
the oscilloscope is 1M
Ω )
Lock the probe until the UNLOCK symbol disappear.
Press the button to degauss and zero set. There will be beeping as success
indication after 6s