122 SSG5000X Programming Guide
the RF output path
Parameter
Type
Float
Parameter
Range
-10 ~ 10
Return
Float
Default
0
Menu
I/Q Control > I/Q Adjustment > Quad Angle Adjustment
Example
:DM:IQADjustment:QSKEW 1
3.8.6.9
I/Q Output
([:SOURce]:DM:IQADjustment:EXTernal[:STATe])
Command
Format
[:SOURce]:DM:IQADjustment:EXTernal[:STATe] ON|OFF|1|0
[:SOURce]:DM:IQADjustment:EXTernal[:STATe]?
Instruction
This command enables or disables the I/Q output
Parameter
Type
Boolean
Parameter
Range
ON|OFF|1|0
Return
Boolean
Default
0
Menu
I/Q Control > I/Q Output
Example
:DM:IQADjustment:EXTernal 1
3.8.6.10
I/Q Output Atten
([:SOURce]:DM:IQADjustment:EXTernal:IQATten)
Command
Format
[:SOURce]:DM:IQADjustment:EXTernal:IQATten <val>
[:SOURce]:DM:IQADjustment:EXTernal:IQATten?
Instruction
This command sets the I/Q output attenuation level (dB).
Parameter
Type
Float
Parameter
Range
0 ~ 20